Calculating The Total Magnification/Field Of View Diameter - Leica S8 APO B User Manual

Table of Contents

Advertisement

Calculating the Total Magnification/Field of View Diameter

Parameters
M O
Magnification of the additional objective
M E
Magnification of eyepiece
z
Magnification changer position
N FOV Field number of the eyepiece. Field numbers are printed on
the eyepieces: 10×/23, 16×/16, 20×/12, 10×/23B, 16×14B,
25×/9.5B, 40×6B.
Magnification in the binocular tube
M TOT VIS = M O × M E × z
or
1.6 × 20 × 4 = 128
Leica S Series
Example
M O
Additional objective 1.6×
M E
20×/12 eyepiece
z
Zoom position 4.0
Calculation example: field of view diameter in the specimen
N FOV
∅ OF:
M O × z
User Manual
12
=
= 1.9 mm
1.6 × 4
48

Advertisement

Table of Contents
loading

Table of Contents