INDEX 1.0 DESCRIPTION page 4 2.0 INTRODUCTION page 6 3.0 UNPACKING AND ASSEMBLY page 6 4.0 USING THE MICROSCOPE page 8 5.0 MAINTENANCE page 12 6.0 TECHNICAL SPECIFICATIONS page 13 7.0 RECOVERY AND RECYCLING page 14 Page 2...
Page 3
Please read this manual carefully before using the equipment. To ensure safe use, the user must read and follow all instructions in this manual. OPTIKA products are designed for safe use in normal operating conditions. The equipment and accessories described in the manual are manufactured and tested according to industry standards for safety instrumentation laboratory.
It is built to high optical and mechanical standards and to withstand daily laboratory use. Optika reminds you that this manual contains important information on safety and maintenance, and that it must therefore be made accessible to the instrument users.
Page 7
3.0 UNPACKING AND ASSEMBLY Insert the observation head onto the top of the analyzer/bertrand lens block and lock its screw. If you have a trinocular microscope and want to use the photo port, insert your adaptor in the hole on the top of the observation head and fixate it with the screw on the side: Insert the eyepieces into the eye tubes.
3.0 UNPACKING AND ASSEMBLY Connect the mains plug into the socket at the base Make sure, before you turn the illumination on, that the voltage selector is set to the mains voltage for your region. The power cord should be used only on network sockets equipped with adequate grounding. Contact a technician to check the state of your electrical system.
Page 9
4.0 USING THE MICROSCOPE Adjust interpupillary distance Hold the right and left parts of the observation head using both hands and adjust the interpupillary distance by turning the two parts until one circle of light can be seen. Focus and diopter adjustment Loosen the focus-stop knob, rotate the coarse focusing knob to bring the slide into focus with a 4X or 10X objective, then lock the focus-stop knob.
Page 10
4.0 USING THE MICROSCOPE Set the numerical aperture Adjust the aperture of the iris diaphragm under the condenser to set the numerical aperture of the illuminator, thus controlling image contrast and resolution. 4.10 Bertrand lens In order to have a conoscopic view of your sample, you can insert a Bertrand lens in the optical path by rotating the disc under the observation head.
Page 11
4.0 USING THE MICROSCOPE 4.11 Polarizer and analyzer filters Insert or remove the lower polarizer as required. It can also be rotated at any angle: The analyzer filter can be inserted by means of its lever under the observation head. By holding this lever, it can laso be rotated at any angle: Page 11...
4.0 USING THE MICROSCOPE 4.12 Retardation filter plates The microscope is provided with these retardation plates: l, l/4 and quartz wedge. You can insert them in the optical path, sliding the plate holder in the slot under the observation head: 5.0 MAINTENANCE Always think about The following environment is required: Indoor temperature: 0-40°C, Maximum relative humidity: 85 % (non condensing)
6.0 TECHNICAL SPECIFICATIONS ILLUMINATION Light source type X-LED with white LED; light intensity control using a knob on left side of the frame. LED power 3.6W, comparable to an halogen bulb 35W. Color temperature: 6300K LED average lifetime approx.: 50.000h. The light exit can be used as a filter holder for additional filters (blue, yellow, frosted).
7.0 RECOVERY AND RECYCLING Art.13 Dlsg 25 july 2005 N°151. “According to directives 2002/95/EC, 2002/96/EC and 2003/108/EC relating to the reduction in the use of hazardous substances in electrical and electronic equipment and waste dispo- sal.” The basket symbol on equipment or on its box indicates that the product at the end of its useful life should be collected separately from other waste.
Page 16
Puig i Pidemunt, nº 28 1º 2ª - (Pol. Ind. Plà d’en Boet) 08302 MATARO (Barcelona) España Tel: +34 937.586.245 +34 937.414.529 New York Microscope Company Inc 100 Lauman Lane, Suite A, Hicksville, New York 11801, USA Tel.: 877.877.7274 - Fax: 516.801.2046 www.microscopeinternational.com - info@nyscopes.com OPTIKA MICROSCOPES - ITALY www.optikamicroscopes.com - info@optikamicroscopes.com...