Philips Lifeline FD100 Instructions For Use Manual page 25

Autoalert help button
Table of Contents

Advertisement

Immunity Test
Electrostatic
Discharge (ESD)
EN/IEC 61000-4-2
Power Frequency
50/60Hz
Magnetic Field
EN/IEC 61000-4-8
Conducted RF
EN/IEC 61000-4-6
Radiated RF
EN/IEC 61000-4-3
NOTE:
Conducted RF Immunity does not apply to an internal battery
operated device.
* P is the max power in watts and D is the recommended separation distance
in meters.
Field strengths from fixed transmitters, as determined by an electromagnetic site
survey, should be less than the compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment containing a transmitter.
EN/IEC 60601
Test Level
6kVContact
8kV Air
3A/m
3 Vrms
150 kHz
to 80 MHz
10 V/m
80 MHz
to 2.5 GHz
Compliance
Electromagnetic
Level
Environment – Guidance
Floors should be wood,
6kV
concrete or ceramic tile.
Contact
If floors are synthetic the
8kV Air
relative humidity should be
at least 30%
3A/m
Power frequency magnetic
fields should be that of
a typical commercial or
hospital environment
not
D=1.17√P
applicable
D=0.35√P
80 to 800 MHz
D=0.70√P
800 MHz to 2.5 GHz
10V/m
Portable and mobile
communications equipment
should be separated
from the AutoAlert by
no less than the distances
calculated/ listed below
D=(3.5/V1)(√P)
D=(3.5/E1)(√P)
80 to 800 MHz
D=(7/Eq)(√P)
800 MHz to 2.5 GHz
For assistance, call 1-800-635-6156
:
*
25

Advertisement

Table of Contents
loading

Table of Contents