Operation Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SSB Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Schuler Scientific SSB Series

  • Page 1 SSB-Series Balances Operation Manual...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS SAFETY PRECAUTIONS……………….……………………………………………………………………………. 1 GETTING STARTED…..…………….………………………………………………………………………………. 1 INSTALLATION & SET-UP……..………………………………………………………….………………………. 2 POWER CONNECTION………….……………………………………………………..…………………………… 3 ANTI-THEFT DEVICE…………………………………………………………………..………….………………… 3 LEVELING..…………………………………………………………………….……………………………………… 3 CALIBRATION…..……………………………………………………….…………………………………………… 4 TARE FUNCTION.……………………………………………………….…………………………………………… 4 GOOD WEIGHING TECHNIQUE…………………………………….…………………………………………… 5 DISPLAY OVERVIEW............…..……………………………………………………… 6 PROGRAMMING INSTRUCTIONS…………………………….………………………………………………… 7 WEIGHING PROGRAMS Basic Weighing Mode.………………………………………………..…………………………………... 7 Multiple Weigh Mode.....……………………………………………….……………………………...
  • Page 3: Safety Precautions

    Connect only accessories and options that are optimally designed for use with your balance. The operator shall be responsible for any modifications to the equipment and for any connection of cables, power supplies or peripheral equipment that are not compatible with SSB Series Balances. Unplug...
  • Page 4: Installation & Setup

    INSTALLATION & SETUP Choose a location that is free from the influences of the following: Drafts: Fans, Heat/Air Duct, Doorways or High Traffic Areas. Vibrations. Extreme Temperature Fluctuations: Sunlight, Ovens, or Environments with Wide Temperature Variations. Voltage Fluctuations. Do Not Share Current with Other Instruments that Draw Voltage in an Inconsistent Manner such as Fluorescent Lights, Centrifuges and the like.
  • Page 5: Power Connection

    AC POWER CONNECTION - When connecting power use the original AC adapter supplied with the balance. Using an unapproved power supply may damage the instrument and void the warranty. - Ensure the power supply is compatible with your instrument AND the local line voltage. - Connect the plug to the instrument.
  • Page 6: Calibration

    CALIBRATION Calibration is required at installation (see Installation & Set-up) and at regular intervals thereafter. Calibrate if the balance is moved to a new location. 1. Allow a 60-minute warm-up period. 2. Press Tare to zero the balance (0.0000g) – if required. 3.
  • Page 7: Good Weighing Technique

    Good Weighing Technique Working with precision balances requires a steady hand and an even, uninterrupted technique. Use forceps or other suitable utensils to place the sample (and sample container) on the weigh pan as fingers are hygroscopic. If the balance has been idle and the draft chamber doors closed for an extended period, perform a number of test measurements before you begin weighing to allow the atmospheric conditions (temperature, RH and barometric pressure) inside the weighing chamber adjust to the ambient temperature outside the chamber.
  • Page 8: Display Overview

    DISPLAY OVERVIEW (1) (3) (5) 00000.000 (6) Key Pad # Function On / Off Key Tare Keys (Right & Left Sides) Calibration Key Mode Key: g, lb, oz, ct, mg or kg, ozt, dwt, gn, Percent (%), Count & Animal Weigh Milligram (mg) Mode (0.01mg, 0.1mg &...
  • Page 9: Programming Instructions

    PROGRAMMING INSTRUCTIONS The SSB-Series offers several Weigh Modes, Balance Settings and Interface Options. Weigh Modes include Gram (g) Milligram (mg) / Kilogram (kg) and Percent Weighing. The SSB-Series also offers adjustable settings for Auto Zero, Baud Rates, Print (data send) Commands, Electronic Filtering and Audible Key Tone functions.
  • Page 10: Balance Setting Programs

    five Initial Sample Setting Option (see below). Follow the Programming Instructions listed above to adjust the Initialize Size that is appropriate to your sample. Once you have selected the Initial Sample Size Press the Mode Key until the COUNT appears on the display. Place the reference pieces on the weigh pan and Press CAL Key to assign the reference sample a value of your parts.
  • Page 11: Data Interface Settings

    3) Audible Key Tone Setting SSB-Series come with an adjustable Audible Key Tone Setting. The factory default setting is on. However, you may enable the Audible Key Tone by following the Programming Instructions listed on page 7. Audible Key Tone Settings Always Off C6 - - 0 Always On...
  • Page 12: Programming Menu Codes - Quick Reference Guide

    PROGRAMMING MENU CODES – QUICK REFERENCE GUIDE Code Function Parts Counting C1 - - 0* Sample Size = 10 C1 - - 1 Sample Size = 20 C1 - - 2 Sample Size = 50 C1 - - 3 Sample Size = 100 C1 - - 4 Sample Size = 1000 Reserved –...
  • Page 13: Care & Maintenance

    CARE & MAINTENANCE ü Unplug the instrument from the power source – wall outlet. ü Wear appropriate safety protection. ü Clean inside hood when appropriate. A damp cloth is recommend for basic cleaning. Do not use solvents on the instrument housing, sub-pan, keypad, power cord and power jack. ü...
  • Page 14: Service & Technical Suypport

    If you have ANY questions or require technical, application or service support please contact Schuler Scientific at 1-800-539-1886. Please note that Schuler Scientific and affiliates will not take back any instrument that has been exposed to biological or hazardous material contamination for replacement, credit, repair or disposal.
  • Page 15: Specifications

    SPECIFICATIONS Model Capacity Readability Repeatability Tare Range SSB-25 0.01mg 0.01mg 0 to 20g SSB-64 0.1mg 0.1mg 0 to 60g SSB-124 120g 0.1mg 0.1mg 0 to 110g SSB-214 210g 0.1mg 0.1mg 0 to 210g SSB-203DS 200g 0.001g 0.001g 0 to 200g SSB-403DS 400g 0.001g...
  • Page 16: Instrument Diagram

    INSTRUMENT DIAGRAMS SSB-25, 64, 124, 214, 203DS, 403DS, 503DS & 1003DS SSB-4002 & 5002 Page 14...
  • Page 17 The status of the information and specifications in this manual is indicated by the date given below. Schuler Scientific reserves the right to make changes to any or all of the specifications, features or design of the instruments at any time and without notice.