3.2.2.4. Z Feedback
All of the parameters for maintaining
Z
feedback are
provided here. The parameters are:
,.
"Gain": Scale factor for the feed,back control signal.
"Proportional": Scale factor for the proportional term in the
PID feedback controller.
"Integra/": Scale factor for the integral term in the PID
feedback controller.
"Derivative": Scale factor for the derivative term in the PID
feedback controller.
"Setpoint": Parameter that controls the magnitude of
interaction between the probe and surface.
Caution: If the
"z
Feedback" parameters are
too high, the
Z
ceramic can begin to oscillate
and potentially cause damage to the ceramic.
. " ",... -,.•.•... 'r·:
~'-'
3.2.2.5. Display
DisI'Ja~j_
There are several options fOf changing the appearance of
"7e~~'ma;:,
>'
'i_~~~E~~.
the images displayed in the "Scan" window. They are:
Right
Image
rz_ERR
"Left Image": The source of the image displayed in the left
f
image window.
I
. . . .
Background
fLi"~;C;;;;-;g~
.
Displa~
Light Shaded
"Right Image": The source of the image displayed in the
right image window.
!
Pallet
Pall"t 1
"Background": When "Line Leveling" is selected, the
background is subtracted from the image one line at
a
time.
"Display": Two types of displays may be selected: "Color
Map" and "Light Shaded."
f'!':.'"'~u;
Y..­
3.2.2.6. Tip Retract
C~nt[Q1
.
After scanning a sample, this function is used to move the
j
.~Qp.
J
r'~~~~~
tip away from the surface. The "Tip Retract" function
Complete
should always be used to assure that the probe does not
I
i
>
I
Start'·
get damaged after scanning.
i'
I
Stop'
3.2.2.7. Control
.j
i
~-
I-E-~i-t
.-,
,
These functions start a scan, stop a scan and exit the
AFM-View software.
t
,
It
I
Load T-APp rc11
4
Z Feedback
t

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for AFM Workshop TT-AFM

Table of Contents