Cleaning

The sling should be washed regularly in water temperature of 140
solution. A soft cloth, dampened with water and a small amount of mild detergent, is all that is
needed to clean the patient lift. The lift can be cleaned with non-abrasive cleaners.

Electric Actuator Replacement

1. Remove the nut, washer and shoulder bolt that secure the electric actuator to the mast
mounting bracket.
2. Rest the boom on your shoulder and remove the nut, bolt, plastic bushing and pinch guard
from the boom mounting bracket.
3. Remove the electric actuator assembly.
4. Reverse steps for installation.

Base Adjustment

The base adjustment should not require any attention
other than:
1. Check the squareness of the legs when in the
CLOSED position.
2. Place a square on the inside of the legs and base to
o
determine the 90
alignment.
3. Adjust the linkage rods until 90
achieved.