Lifter Maintenance - Invacare RPS 440ee Instruction Manual

Table of Contents

Advertisement

Lifter Maintenance

Safety Inspections
A person who is suitably and properly qualified and well acquainted with the design, use and
care of the hoist should perform periodic inspections.
Inspections should be carried out both routinely (at least as often as specified) and, as necessary.
All defects and damages that have lead to corrective action should be noted in a log book
(see page 26) and signed and dated by the inspector. Defects and resulting actions should be
reported in writing to your supplier.
If inspection reveals that the safety of the hoist is jeopardised in any way, the Owner should
be notified immediately, the hoist should be taken out of service and not used again until the
problem has been rectified. Contact the supplier to arrange servicing.
A Guide to Maintenance Procedures
A regular visual check of the hoist and attachments is essential. Invacare Australia Pty Ltd
permits servicing only by an Authorised Service Agent in accordance with the following schedule
of Service Types.
Item
PRE-DELIVERY
Confirm all components are included
Inspect parts for possible damage in transit
COMPLETE LIFTER
General inspection of Lifter. Visually inspect the patient lifter for
external damage or wear. Scratches and chips.
Retard rusting by applying spray paint (Dulux 'Special White' Code:
81880)
BASE AND CASTORS
Inspect for missing hardware
Base opens/closes with ease.
Inspect castors and axle bolts for tightness. Make sure tightly attached
to legs.
Inspect castors for smooth swivel and roll
Check castor brakes operate efficiently
Overhaul castors and repack with grease
Check tightness of bolt in leg bushes
Check linkages are secure in leg adjuster ball joints
MAST
Mast MUST be securely assembled to boom
Inspect for bends or deflections.
Check sling hooks for wear or deflection.
ELECTRIC ACTUATOR ASSEMBLY
Check hand control and cycle to ensure smooth quiet operation. Lift
adjustment on all Lifters. Width also for those with electric leg adjustment.
Emergency lowering function (electric and mechanical)
Check for wear or deterioration. (IF DAMAGED, RETURN TO FACTORY)
Inspect hardware on mast and boom.
Remove dust and dirt from actuator with cloth
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare
®
RPS 440ee User's Manual
20
is a registered trademark and Yes, you can.™ is a trademark of Invacare
Pre-Delivery
Daily
Minor Service
(After every
3 Months of
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
®
Corporation. 2008 Invacare Asia Pacific
Major Service
(After every
6 Months of
Service)
Service)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

Hide quick links: