3 series blood pressure monitor

Advertisement

INSTRUCTION MANUAL
3 series
Blood Pressure Monitor
Model BP710CANN

Advertisement

Table of Contents
loading

Summary of Contents for Omron BP710CANN

  • Page 1 INSTRUCTION MANUAL 3 series Blood Pressure Monitor Model BP710CANN...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Before using the Monitor Introduction ....................3 Important Safety Information..............4 Operating Instructions Know Your Device..................7 Display symbols ............... 9 Before Taking a Measurement ..........10 Preparation ....................11 Battery Installation..............11 Using the Device..................12 Applying the Arm Cuff ............12 How to Sit Correctly..............
  • Page 3: Introduction

    INTRODUCTION ® Thank you for purchasing the OMRON BP710CANN Blood Pressure Monitor. Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading.
  • Page 4: Important Safety Information

    IMPORTANT SAFETY INFORMATION Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.
  • Page 5 IMPORTANT SAFETY INFORMATION (General Usage) Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous. Consult your physician before using the device for any of the following conditions: •If you have had a mastectomy. •Do not take measurements more than necessary. It may cause bruising due to blood flow interference.
  • Page 6 IMPORTANT SAFETY INFORMATION (Battery Usage) Do not insert the batteries with their polarities incorrectly aligned. Use only 4 “AA” alkaline or manganese batteries with this device. Do not use other types of batteries. Do not use new and used batteries together. Remove the batteries if the device will not be used for three months or more.
  • Page 7: Know Your Device

    KNOW YOUR DEVICE Contents: Monitor, arm cuff, instruction manual, quick start guide Monitor: A. Display D. Battery compartment B. START/STOP button E. Air jack C. Memory button Arm cuff: F. Arm cuff (Arm circumference 9” - 17” (22 - 42 cm)) G.
  • Page 8 KNOW YOUR DEVICE Display: Memory symbol N. Irregular heartbeat symbol J. Systolic blood pressure O. Movement error symbol K. Diastolic blood pressure P. Pulse display / Memory number L. Low battery symbol Q. Deflation symbol M. Heartbeat symbol (Flashes during measurement.)
  • Page 9: Display Symbols

    KNOW YOUR DEVICE Display symbols Irregular Heartbeat Symbol ( Normal Heartbeat When the monitor detects an irregular Pulse rhythm two or more times during the measurement, the irregular heartbeat Blood pressure symbol will appear on the display with the Irregular Heartbeat measurement values.
  • Page 10: Before Taking A Measurement

    KNOW YOUR DEVICE Before Taking a Measurement To help ensure an accurate reading, follow these directions: 1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement.
  • Page 11: Preparation

    PREPARATION Battery Installation Remove the battery cover. Insert 4 “AA” batteries as indicated in the battery compartment. Replace the battery cover. Notes: •When the low battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same time. Long life alkaline batteries are recommended.
  • Page 12: Using The Device

    USING THE DEVICE Applying the Arm Cuff Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm. Do not place the arm cuff over thick clothes. Insert the air plug into the air jack securely. Wrap the arm cuff firmly in place around your left upper arm.
  • Page 13 • The blood pressure can differ between the right arm and the left arm, and the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician...
  • Page 14: How To Sit Correctly

    USING THE DEVICE How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement. • Sit on a chair with your feet flat on the floor.
  • Page 15 USING THE DEVICE Press the START/STOP button. The arm cuff will start to inflate automatically. START INFLATE DEFLATE COMPLETE If your systolic pressure is more than 210 mmHg After the arm cuff starts to inflate, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
  • Page 16 USING THE DEVICE Remove the arm cuff. Press the START/STOP button to turn the monitor off. The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes. Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement.
  • Page 17: Using The Memory Function

    USING THE DEVICE Using the Memory Function The monitor automatically stores the results up to 14 sets. Note: If the memory is full, the monitor will delete the oldest value. To View the Measurement Values Stored in Memory Press the button.
  • Page 18 USING THE DEVICE To Delete All the Values Stored in Memory Press the button, while the memory symbol appears. While holding the button down, press the START/STOP button for more than 3 seconds. Note: You cannot partially delete the values stored in the memory.
  • Page 19: Error Messages And Troubleshooting

    ERROR MESSAGES AND TROUBLESHOOTING Error Messages Error Display Cause Solution Remove the arm cuff. Wait 2 - 3 minutes and then take another Irregular heartbeats are measurement. detected. Repeat the steps in Taking a Measurement. If this error continues to appear, contact your physician. Movement during Carefully read and repeat the steps measurement.
  • Page 20 ERROR MESSAGES AND TROUBLESHOOTING Error Display Cause Solution Repeat measurement. Remain still and do not talk during measurement. Refer to Taking a Measurement. Movement during measurement and the If “E2” appears repeatedly, inflate arm cuff has not been the arm cuff manually until it is 30 to inflated sufficiently.
  • Page 21: Troubleshooting

    ERROR MESSAGES AND TROUBLESHOOTING Troubleshooting Problem Cause and Solution Replace all batteries with new ones. No power. Check the battery installation for proper placement of the battery polarities. No display appears on the monitor. Refer to Battery Installation. Blood pressure varies constantly. Many factors including stress, time of day, Measurement values appear too high and how you wrap the cuff, may affect...
  • Page 22: Maintenance And Storage

    MAINTENANCE AND STORAGE Maintenance To protect your device from damage, please observe the following: • Store the device and the components in a clean, safe location. • Do not use any abrasive or volatile cleaners. • Do not wash the device and any components or immerse them in water.
  • Page 23: Storage

    MAINTENANCE AND STORAGE Storage Unplug the air plug from the air jack. Gently fold the air tube into the arm cuff. Note: Do not bend or crease the air tube excessively. Do not store the device in the following situations: •If the device is wet.
  • Page 24: Specifications

    SPECIFICATIONS BP710CANN HEM-7121-CA Model Display LCD digital display Measurement range Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats / min. Pressure: ±3 mmHg or 2% of reading Accuracy Pulse: ± 5% of display reading Inflation Fuzzy-logic controlled by electric pump...
  • Page 25 SPECIFICATIONS Notes: • These specifications are subject to change without notice. • In the clinical validation study, the 5th phase was used on 85 subjects for determination of diastolic blood pressure. • This device has not been validated for use on pregnant patients.
  • Page 26: Fcc Statement

    FCC STATEMENT FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
  • Page 27: Limited Warranty

    PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY. OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR DAMAGES.
  • Page 28 LIMITED WARRANTY This warranty provides you with specific legal rights, and you may have other rights that vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you. FOR CUSTOMER SERVICE Visit our web site at: www.omronhealthcare.ca Call toll free:...
  • Page 29: Guidance And Manufacturer's Declaration

    EMC and needs to be installed and put into service according to the EMC information provided in this documentations. • The Essential Performance of the BP710CANN is to measure a blood pressure and a pulse rate and using the memory function.
  • Page 30 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP710CANN is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP710CANN should assure that it is used in such environment. Emissions test Compliance...
  • Page 31 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP710CANN is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP710CANN should assure that it is used in such environment. Electromagnetic Immunity test...
  • Page 32 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP710CANN is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP710CANN should assure that it is used in such environment. IEC 60601 Compliance...
  • Page 33 RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the OMRON BP710CANN is used exceeds the applicable RF compliance level above, the OMRON BP710CANN should be observed to verify normal operation.
  • Page 34 Manufactured for: OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN Distributed by: OMRON HEALTHCARE, INC. 1925 West Field Court Lake Forest, IL 60045 U.S.A. www.omronhealthcare.ca © 2014 OMRON HEALTHCARE, INC. Made in China 2298782-2A...