Ordinary Location Certification Csa; Hazardous Locations Certificates - Emerson CSI 9420 Reference Manual

Wireless vibration transmitter
Hide thumbs Also See for CSI 9420:
Table of Contents

Advertisement

CSI 9420
Ordinary Location
Certification CSA
HAZARDOUS
LOCATIONS
CERTIFICATES
B-2
2.4 GHz:
FCC ID: LW2RM2510
IC ID: 2731A-RM2510
As standard, the transmitter has been examined and tested to determine that
the design meets basic electrical, mechanical, and fire protection
requirements by CSA, a nationally recognized testing laboratory (NRTL) as
accredited by the Federal Occupational Safety and Health Administration
(OSHA).
CE mark
The 2.4 GHz version of the device has been tested and complies with all
relevant directives required for CE marking. The 900 MHz verson of the device
was not evaluated for CE.
Electro Magnetic Compatibility (EMC) (2004/108/EC)
All Models: EN 61326-1, 61326-2-3: 2006
CSA - Canadian Standards Association
CAN/CSA-C22.2 No. 61010-1-04 - Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use, Part 1: General
Requirements
ISA S82.02.01 2nd (IEC 61010-1 Mod) - Safety Standards for Electrical and
Electronic Test, Measuring, Controlling and Related Equipment - General
Requirements
UL 61010-1 2nd - Safety Requirements for Electrical Equipment for
Measurements, Control , and Laboratory Use, Part 1: General Requirements
North American Certifications
Factory Mutual (FM) Approvals
FM Non-incendive
Non-incendive for Class I, Division 2, Groups A, B, C, and D.
Temperature Codes T4 (-40 °C < Ta < 85 °C)
Ambient temperature limits: -34 to 80 °C
For use with Rosemount power module pack P/N 753-9220-XXXX (MHM
P/N 89002) only.
Enclosure Type 4X / IP66
Zone Marking: Class I, Zone 2, Group llC
T5 (-20 °C < Ta < 80 °C) with LCD
Reference Manual
MHM-97408.5
April 2010