SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements:
Rated Current Input:
Sound Level:
Altitude:
*Oxygen Output
Concentration Levels:
Maximum Outlet Pressure:
Flow Range:
Low Flow Alarm:
Average Power
Consumption:
Pressure Relief Mechanism
Operational at:
Change in maximum recom-
mended flow when back pres-
sure of 7kPa is applied:
Filters:
Safety System:
Part No 1143482
SECTION 4—TYPICAL PRODUCT PARAMETERS
120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
3 A
44 ± 2 dBA Average
Up to 6,000 ft (1828 meters) above sea level
without degradation of concentration levels.
From 6,000 ft (1828 meters) to 13,129 ft
(4000 meters) below 90% efficiency.
87% to 95.6% for flows ranging at 0.5 to 5 L/min
5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
0.5 to 5 L/min
For flowrates less than 1 L/min, we recom-
mend the use of the Invacare Pediatric Flow-
meter IRCPF16 Accessory.
0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No
Accessories Connected).
280 W (Operating @ 3 L/min)
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
0.7 L/min
Cabinet, Outlet HEPA and Compressor Inlet
Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor shut-
down. Low Pressure Alarm w/compressor
shutdown. Battery Free Power Loss Alarm.
SensO
Flow Alarm.
Oxygen System (SensO
2
13
Model) Low
2
Perfecto
Series
2