SECTION 4—TYPICAL PRODUCT PARAMETERS
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements:
Rated Current Input:
Sound Level:
Altitude:
*Oxygen Output
Concentration Levels:
Maximum Outlet Pres-
sure:
Flow Range:
Low Flow Alarm:
Average Power
Consumption:
Pressure Relief Mecha-
nism Operational at:
Perfecto
Series
2
120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
3 A
43 ± 2 dBA Average for Perfecto
Perfecto
V Model
2
40 ± 2 dBA Average for Perfecto
All Perfecto
Models: Up to 8,000 ft (2438
2
meters) above sea level without degredation of
concentration levels. From 8,000 ft to 13,129
(4000 meters) below 90% efficiency.
All Perfecto
Models while filling a
2
HomeFill System: Up to 6,000 ft (1828 meters)
above sea level without degradation of
concentration levels. From 6,000 ft (1828 meters)
to 13,129 ft (4000 meters) below 90% efficiency.
87% to 95.6% for flows ranging at 0.5 to 5 L/min
5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
0.5 to 5 L/min
For flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.
0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No Accessories
Connected).
Perfecto
and Perfecto
2
(Operating @ 3 L/min)
Perfecto
V Model: 325 W
2
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
12
Models and
2
W Models
2
W Models: 280 W
2
Part No 1143482