Omron IntelliSense HEM-790ITCAN Instruction Manual
Omron IntelliSense HEM-790ITCAN Instruction Manual

Omron IntelliSense HEM-790ITCAN Instruction Manual

With comfit cuff

Advertisement

Quick Links

INSTRUCTION MANUAL
Automatic Blood Pressure Monitor
with ComFit
Cuff
TM
HEM-790ITCAN
Model

Advertisement

Table of Contents
loading

Summary of Contents for Omron IntelliSense HEM-790ITCAN

  • Page 1 INSTRUCTION MANUAL Automatic Blood Pressure Monitor with ComFit Cuff HEM-790ITCAN Model...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Before Using the Monitor Introduction ..........4 Safety Information .
  • Page 3 TABLE OF CONTENTS Taking a Measurement .........28 Using the Guest Mode .
  • Page 4: Introduction

    TruRead allows for three consecutive measurements with timed intervals for an average reading. Omron Health Management Software included with this unit allows you to view, process and print data about blood pressure measured by the HEM-790ITCAN. The HEM-790ITCAN comes with the following components: •...
  • Page 5: Safety Information

    SAFETY INFORMATION To assure the correct use of the product basic safety measures should always be followed including the warnings and cautions listed in this instruction manual. SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL Indicates a potentially hazardous situation which, if not avoided, could result in death or WARNING serious injury.
  • Page 6 Do not use a cellular phone near the device. It may result in an operational failure. Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.
  • Page 7: Risk Of Electrical Shock

    Do not submerge the device or any of the components in water . Store the device and the components in a clean, safe location. Changes or modification not approved by Omron Healthcare will void the user warranty. Do not disassemble or attempt to repair the...
  • Page 8: Before Taking A Measurement

    BEFORE TAKING A MEASUREMENT To ensure a reliable reading follow these recommendations: 1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement. 2. Stress raises blood pressure. Avoid taking measurements during stressful times.
  • Page 9: Know Your Unit

    KNOW YOUR UNIT Main Unit: Display USER ID Selection Switch Battery Compartment Start/Stop Button USB Port Memory ( ) Button Air Jack Morning/Evening Average AC Adapter Jack ) Buttons SET Button...
  • Page 10 KNOW YOUR UNIT Components: Housing Air Tube Cord AC Adapter Air Plug (Connects to Air Jack) Arm Cuff Four “AA” Batteries Storage Case USB Cable CD-ROM INSTRUCTION MANUAL Automatic Blood Pressure Monitor w th ComFit Cuff HEM 790ITCAN Model Instruction Manual...
  • Page 11: Unit Display

    UNIT DISPLAY Systolic Blood Pressure USER ID Symbol (A or B) Diastolic Blood Pressure Morning Hypertension Symbol Heartbeat Symbol Morning Average Symbol Movement Symbol Evening Average Symbol Irregular Heartbeat Symbol Deflation Symbol Average Value Symbol Date/Time Display Memory Symbol Low Battery Indicator Pulse Display Week Display TruRead...
  • Page 12: Display Symbols

    DISPLAY SYMBOLS IRREGULAR HEARTBEAT SYMBOL ( When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the measurement values. An irregular heartbeat rhythm is defined as Normal Hear e at a rhythm that is more than 25% slower or...
  • Page 13: Morning Hypertension Symbol

    Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider. Omron Healthcare recommends you contact your physician if the MORNING HYPERTENSION SYMBOL and/or the HEARTBEAT SYMBOL appear on the display with your measurement values.
  • Page 14: Heartbeat Symbol

    DISPLAY SYMBOLS HEARTBEAT SYMBOL ( TAKING A MEASUREMENT The Heartbeat Symbol flashes on the display during the measurement. The Heartbeat Symbol flashes at every heartbeat. When the measurement is complete, the Heartbeat Symbol flashes on the display with your blood pressure and pulse rate if the reading is above 135 for the Systolic Blood Pressure value and/or 85 for the Diastolic Blood Pressure value.
  • Page 15: Battery Installation

    BATTERY INSTALLATION Battery Cover 1. Press the ▼ indicator on the battery cover and slide the cover off in the direction of the arrow. 2. Install 4 “AA” size batteries so the + (positive) and - (negative) polarities match the polarities of the battery compartment Å...
  • Page 16: Battery Replacement

    BATTERY INSTALLATION BATTERY REPLACEMENT Low Battery Indicator When the Low Battery Indicator appears on the display screen turn the monitor off and remove all the batteries. Replace with four new batteries at the same time. Longlife alkaline batteries are recommended. CAUTION If battery fluid should get in your eyes, immediately rinse with plenty of clean water.
  • Page 17: Using The Ac Adapter

    USING THE AC ADAPTER Use only the authorized Omron AC Adapter with this monitor. We recommend you install the batteries even if the AC Adapter is used. If no batteries are installed, you may need to reset the date and time if the AC Adapter is disconnected.
  • Page 18 The power unit (AC Adapter) is intended to be correctly oriented in a vertical or floor mount position. NOTES: • Use only the authorized Omron AC Adapter that came with this monitor: • We recommend you install the batteries even if the AC Adapter is used.
  • Page 19 USING THE AC ADAPTER TO CONNECT THE AC ADAPTER Insert the AC Adapter Plug into Housing the AC Adapter Jack on the back side of the monitor as illustrated. Plug the AC Adapter into a 120V AC outlet (60 cycles). Cord TO DISCONNECT THE AC ADAPTER Hold the Housing to remove the AC Adapter from the electrical...
  • Page 20: Setting The Date And Time

    SETTING THE DATE AND TIME Set the monitor to the current date and time before taking a measurement for the first time. If the date and time are not set correctly , the measurement values stored in the memory and the average values may not be correct.
  • Page 21 SETTING THE DATE AND TIME 2. SETTING THE YEAR The year can be set between 2007 and 2030. When the display reaches 2030, it will return to 2007. Press the Memory button ( ) to advance by increments of one year. Press the SET button to set the current year.
  • Page 22 SETTING THE DATE AND TIME 4. SETTING THE DAY Press the Memory button ( ) to advance by increments of one day . Press the SET button to set the current day. The hour flashes on the display. 5. SETTING THE HOUR The time is set using AM or PM.
  • Page 23 SETTING THE DATE AND TIME RESETTING THE DATE AND TIME To adjust the date or time after it has been setup when you first use the unit. Start with the monitor in the of f position. 1. Press and hold the SET button. The TruRead symbol and the current setting (on or oFF)
  • Page 24: Applying The Arm Cuff

    APPLYING THE ARM CUFF CAUTION Read all of the information in the instruction manual and any other literature in the box before operating the unit. CAUTION This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions.
  • Page 25 APPLYING THE ARM CUFF 3. Sit in a chair with your feet flat on the floor . Place your left arm on a table so the cuf f is level with your heart. Thumb Grip 4. Hold the grip on the cuf f securely with your right hand, placing your thumb on the thumb grip.
  • Page 26 APPLYING THE ARM CUFF 6. Apply the cuff to your left upper arm so the blue strip is on the inside of your arm and aligned with your middle finger . The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2"...
  • Page 27: Applying The Cuff On The Right Arm

    APPLYING THE ARM CUFF APPLYING THE CUFF ON THE RIGHT ARM When taking a measurement using the right arm use this instruction for Step 6 on page 25. Apply the cuff to your right upper arm so the thumb grip is centered on the inside of your inner arm.
  • Page 28: Taking A Measurement

    TAKING A MEASUREMENT The monitor is designed to take measurements and store the measurement values in the memory for two people using USER ID A and USER ID B. The monitor can also be used to take a single measurement for other persons using the GUEST Mode.
  • Page 29: Selecting The User Id

    TAKING A MEASUREMENT 3. Inflation stops and the measurement is started. When the measurement is completed, your blood pressure and pulse rate appear on the display. SELECTING THE USER ID Select your USER ID. Always use the same USER ID when taking a measurement.
  • Page 30: Selecting The Truread Tm Mode

    TAKING A MEASUREMENT SELECTING THE TRUREAD MODE 1. Select your USER ID (A or B). 2. Press and hold the SET button. The TruRead symbol and the oFF setting appears on the display. 3. Press the Memory button ( to select on. NOTE: Press the Memory button to alternate between on and oFF.
  • Page 31 TAKING A MEASUREMENT 4. Press the SET button to confirm the change. The number of seconds between each measurement is displayed. NOTE: The default value between each measurement is 60 seconds. 5. Press the MEMORY button ( ) to change the interval of time between each measurement.
  • Page 32: Using The Single Mode

    TAKING A MEASUREMENT 7. If the date and time are correct, press the START/STOP button to turn the monitor off. To change the date and time refer to page 19, SETTING THE DATE AND TIME. CAUTION Do not use a cellular phone near the device. It may result in an operational failure.
  • Page 33 TAKING A MEASUREMENT 2. Press the START/STOP button. All symbols appear on the display. The USER ID Symbol you selected flashes on the display. The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically determines your ideal inflation level. This monitor detects the pulse during inflation.
  • Page 34 TAKING A MEASUREMENT 4. When the measurement is complete, the arm cuf f completely deflates. Your blood pressure and pulse rate are displayed. 5. Press the START/STOP button to turn the monitor off. NOTE: The monitor will automatically turn off after five minutes.
  • Page 35: Using The Truread Tm Mode

    TAKING A MEASUREMENT USING THE TRUREAD MODE 1. Select your USER ID (A or B). 2. Press the START/STOP button. All symbols appear on the display. The USER ID Symbol you selected flashes on the display. The cuff starts to inflate automatically. NOTE: To stop the inflation or measurement, press the ST ART/STOP button.
  • Page 36 TAKING A MEASUREMENT The second measurement is complete, waiting to start the next measurement. The consecutive measurements are complete. The average for three measurements appears on the display. 5. Press the SET button to display the blood pressure and pulse rate for the first measurement.
  • Page 37: Special Conditions

    TAKING A MEASUREMENT SPECIAL CONDITIONS CAUTION Inflating to a higher pressure than necessary may result in bruising where the cuff is applied. If your systolic pressure is known to be more than 220 mmHg, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
  • Page 38: Using The Memory Function

    USING THE MEMORY FUNCTION The monitor is designed to store the blood pressure and the pulse rate in the memory for two people (USER A and USER B) every time a measurement is completed. The monitor automatically stores up to 84 sets of measurement values (blood pressure and pulse rate) for each user (A and B).
  • Page 39 USING THE MEMORY FUNCTION TO DISPLAY THE MEASUREMENT VALUES 1. Select your USER ID (A or B). 2. Press the Memory button ( The Average Symbol ( ) appears on the display when the average reading is shown.
  • Page 40: To Display The Measurement Values

    USING THE MEMORY FUNCTION 3. Press the Memory button ( ) to display the most recent set of measurement values on the screen. The values are displayed from the most recent to the oldest. NOTE: The date and time are alternately displayed with the measurement values.
  • Page 41: Morning And Evening Averages

    USING THE MEMORY FUNCTION MORNING AND EVENING AVERAGES The monitor calculates and displays a weekly average for measurements taken in the morning ( ) and evening ( ). The monitor stores 8 weeks of Morning Averages and 8 weeks of Evening Averages for each user (A and B).
  • Page 42: To Display Morning And Evening Averages

    USING THE MEMORY FUNCTION TO DISPLAY MORNING AND EVENING AVERAGES 1. Select your USER ID (A or B). 2. Press the Morning Average button ) or the Evening Average button ( The average for the current week “THIS WEEK” appears on the display.
  • Page 43: Display Combinations

    USING THE MEMORY FUNCTION DISPLAY COMBINATIONS In addition to the Morning Average and Evening Average Symbols, the monitor may also display the Morning Hypertension Symbol if the morning average for that week is above the CHS guidelines. (See page 13 for details.) Depending on your measurement results, these may be displayed in the following combinations.
  • Page 44: To Delete All Values Stored In The Memory

    USING THE MEMORY FUNCTION The Morning Hypertension Symbol appears if the weekly average for morning measurements is above 135/85. In this case the Morning Hypertension Symbol is displayed when the Evening Average is displayed, regardless of the values for the Evening Average.
  • Page 45: Installing The Software

    INSTALLING THE SOFTWARE Omron Health Management Software Omron Health Management Software allows you to view , process and print data about blood pressure and walking measured by Omron’ s blood pressure monitor and pedometer. NOTE: This software is exclusive to Omron’s blood pressure monitors and pedometers.
  • Page 46 INSTALLING THE SOFTWARE OPERATING ENVIRONMENT REQUIREMENTS The following operating environments are required to run this software. Microsoft Windows 2000 Professional ® ® Supported OS: Microsoft Windows XP Home Edition ® ® Microsoft Windows XP Professional Edition ® ® 600 MHz or higher, Pentium compatible CPU, PC/AT compatible machine 128MB or more Memory capacity:...
  • Page 47: Management Software

    • If you use the included CD-ROM, you can stop reading this NOTE because its version is 1.20 or higher. • If you use the CD-ROM included with an Omron pedometer , please confirm the version. If the version is less than 1.20, update it by using the included CD-ROM or by going to http://software.omronhealthcare.com...
  • Page 48 2-1. Backup your data file 1) Open the [System] folder inside the folder in which you installed the Omron Health Management Software. 2) Copy and paste the [omron mdb] file to the desktop or a storage device. 2-2. Uninstall Omron Health Management Software 1) Select [All programs] or [Program] from [Start] on the PC.
  • Page 49 INSTALLING THE SOFTWARE INSTALLING THE SOFTWARE • To install Omron Health Management Software onto your PC, you must have administrative privileges. • Close all applications currently running before you install the software. • When using anti-virus software, disable anti-virus software before installation.
  • Page 50 INSTALLING THE SOFTWARE 2. The installation screen of MDAC may appear. Click [OK]. If this screen does not appear, skip to Step 5. 3. Select the checkbox [I accept …], and click [Next].
  • Page 51 INSTALLING THE SOFTWARE 4. Click [Finish]. NOTES: • This screen appears when your PC must be restarted. Select the checkbox [Let setup restart the system now], and click [Finish]. Your PC will restart. After restart, start the CD-ROM. (Refer to Step 1 for how to start the CD-ROM.) •...
  • Page 52 INSTALLING THE SOFTWARE 5. The installation screen of the .NET Framework1.1 may appear. Click [Yes]. NOTE: If this screen does not appear, skip to Step 8. 6. Microsoft .NET Framework 1.1 License Agreement appears. Read the License Agreement thoroughly and if you agree with it, select [I agree] and click [Install].
  • Page 53 INSTALLING THE SOFTWARE 8. The software installation guide screen of Omron Health Management Software appears. Click [Next >] to continue installation. 9. The License Agreement of Omron Health Management Software appears. Read the License Agreement thoroughly and if you agree with it, select [I accept the terms..] and click [Next].
  • Page 54 INSTALLING THE SOFTWARE 10.The installation folder selection screen appears. Check the installation destination folder, and click [Next >]. To change the installation destination, click [Change...] and select a folder for installation. 11.The installation confirmation screen appears. To continue installation, click [Install].
  • Page 55 PC must be restarted. Click [Yes] to restart your PC. 13.The Omron Health Management Software shortcut icon is created on the desktop. 14.Remove the CD-ROM disc from the CD drive. 15.Enable anti-virus software, if it had been disabled before you...
  • Page 56: Using The Software

    [Omron Health Management Software Users Manual.pdf] shortcut on desktop, or click [Help] in this software after starting up. 1. Double-click the [Omron Health Management Software] shortcut icon on the desktop. If no icon appears on the desktop: From the [Start]...
  • Page 57 3. In the New User Registration box, enter your name (e.g., Chuck) and click [OK]. NOTES: • Omron Health Management Software can manage multiple users’ data. Data is controlled by user name. • The number of users who can register for this software varies depending on the amount of free disc space in your PC.
  • Page 58: Transferring Data To A Personal Computer

    TRANSFERRING DATA TO A PERSONAL COMPUTER Use only the authorized Omron USB cable, Model HHX-CABLE- USB2, included with this unit. CAUTION Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.
  • Page 59 TRANSFERRING DATA TO A PERSONAL COMPUTER 4. Click [Download Data] icon to import the data. 5. The Confirm dialog appears. Make sure your cable is connected then click [Yes]. 6. Select the blood pressure monitor being used and click [OK]. 7.
  • Page 60: Software Trouble Shooting Tips

    Windows XP Management Software currently installed in operating system is onto a PC on which the your PC is neither required to install Omron corresponding operating Windows 2000 nor Health Management system is running as listed Windows XP. Software onto your PC.
  • Page 61: Care And Maintenance

    CARE AND MAINTENANCE To keep your digital blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed below: Keep the monitor in the storage case when not in use. Make sure the AC adapter is placed under the main unit so that it does not damage the display.
  • Page 62 CARE AND MAINTENANCE Use the unit consistent with the instruction provided in this manual. CAUTION Changes or modification not approved by Omron Healthcare will void the user warranty. Do not disassemble or attempt to repair the unit or components. CAUTION Use only Omron authorized parts and accessories.
  • Page 63: Error Indicators And Troubleshooting Tips

    ERROR INDICATORS AND TROUBLESHOOTING TIPS ERROR INDICATORS SYMBOL CAUSE CORRECTION Cuff under-inflated. Cuff not applied Single Mode correctly. Remove the arm cuff. Read, “Taking a Measurement”. Take another measurement. TruRead Mode Cuff over-inflated The monitor will repeat the process up to 5 times. Monitor could not detect pulse wave.
  • Page 64: Troubleshooting Tips

    ERROR INDICATORS AND TROUBLESHOOTING TIPS TROUBLESHOOTING TIPS PROBLEM CAUSES AND SOLUTIONS No power. Replace all four batteries with new ones. No display appears on the unit. Check the battery installation for proper placement of the battery polarities. Blood pressure varies constantly. Many factors including stress, time of day, and how you wrap the cuf f, Measurement values appear...
  • Page 65: Fcc Statement

    FCC STATEMENT NOTE: POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
  • Page 66: Warranty

    APPLICABLE WRITTEN WARRANTY ABOVE. Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR DAMAGES.
  • Page 67: Specifications

    SPECIFICATIONS Model: HEM-790ITCAN (HEM-7080-ITCAN) Display: LCD Digital Display Measurement Range: Pressure: 0 to 299 mmHg, Pulse: 40-180/min. Accuracy/Calibration: Pressure: ±3mmHg or 2% of reading Pulse: ±5% of reading Inflation: Automatic by electric pump Deflation: Automatic pressure release valve Rapid Air Release:...
  • Page 68 Made in China OMRON HEALTHCARE, INC. 1200 Lakeside Drive Bannockburn, Illinois 60015 www.omronhealthcare.com Copyright © 2009 Omron Healthcare, Inc...

Table of Contents