Oxygen Sensor - Invacare IRC5PO2 Service Manual

Oxygen concentrator homefill compatible sensored and non-sensored
Hide thumbs Also See for IRC5PO2:
Table of Contents

Advertisement

SECTION 24—ALARM TEST
Location of In-Line
Connector
In-Line Connector
FIGURE 24.2 High Pressure Test

Oxygen Sensor

NOTE: For this procedure, refer to FIGURE 24.3 on page 83.
Alarm sequence occurs when concentration levels fall below a preset value, typically 73% 
± 3% or 85% ± 2%. This failure indicates a catastrophic failure of any component 
mechanical or electrical. 
1. Turn unit on ( ). Set output flow at 5 L/min.
2. Monitor the O
 level.
2
3. With the O
 level greater than 85% after five minutes, the GREEN panel indicator light 
2
will remain illuminated.