Agilent Technologies E1465A User Manual page 87

Relay matrix switch modules
Table of Contents

Advertisement

Reading the
Status/Control Register
Writing to the
Status/Control Register
Appendix B
For Status/Control register reads, three bits are defined as follows.
MODID (bit 14):
MODID (module ID) and a 1 indicates the module has not been
selected. For example, if an E1466A matrix module is not busy
(bit 7 = 1) and the interrupt is enabled (bit 6 = 0), a read of the
Status/Control Register (base + 04
Module ID (bits 10 - 13):
the module configuration (E1465A/66A/67A determined by the
terminal module attached).
Busy (bit 7)
about 7 ms execution time during which time the matrix module
is busy. Bit 7 of this register is used to inform the user of a busy
condition.
Enable (bit 6)
generated after a channel has been closed can be disabled. Bit 6
of this register is used to inform the user of the interrupt status.
You can only write to bits 0 and 6 of the Status/Control Register.
Enable (bit 6)
of the module.
Soft Reset (bit 0):
module. To reset each relay in register-based programming, you
must write all 0s to all 16 banks to open all relays.
NOTE
When writing to the registers it is necessary to write "0" to bit 0 after the
reset has been performed before any other commands can be programmed
and executed. SCPI commands take care of this automatically.
0 indicates the module has been selected by
The following bit representations determine
Model/Bits
(13)
E1465A
1
E1466A
0
E1467A
0
: 0 indicates the module is busy. Each relay requires
: 0 indicates the interrupt is enabled. The interrupt
: Writing a "1" to this bit disables the interrupt function
Writing a "1" to this bit does not soft reset the
Register-Based Programming 87
) returns DBBF.
16
(12)
(11)
0
0
1
1
1
0
(10)
1
0
1