E1465A Crosstalk Between Channels
Specifications are for 16 x 16 matrix, for Z(load) = Z(source) = 50
crosspoint closed per row or column. Typical is defined as the worst crosspoint test result from one or two matrix
modules.
Within a Card (worst path)
Closed Path to Closed Path (typical)
Open Row to Open Row (typical)
Open Row to Open Column (typical)
Open Column to Open Column (typical)
Module to Module (Represents 16 x 32 Configuration)*
Closed Path to Closed Path (typical)
Open Row to Open Row (typical)
Open Row to Open Column (typical)
Open Column to Open Column (typical)
E1466A Crosstalk Between Channels
Specifications are for 4 x 64 matrix, for Z(load) = Z(source) = 50
crosspoint closed per row or column. Typical is defined as the worst crosspoint test result from one or two matrix
modules.
Within a Card (worst path)
Closed Path to Closed Path (typical)
Open Row to Open Row (typical)
Open Row to Open Column (typical)
Open Column to Open Column (typical)
Module to Module (Represents 4 x 128 Configuration)*
Closed Path to Closed Path (typical)
Open Row to Open Row (typical)
Open Row to Open Column (typical)
Open Column to Open Column (typical)
E1467A Crosstalk Between Channels
Specifications are for 8 x 32 matrix, for Z(load) = Z(source) = 50
crosspoint closed per row or column. Typical is defined as the worst crosspoint test result from one or two matrix
modules.
Within a Card (worst path)
Closed Path to Closed Path (typical)
Open Row to Open Row (typical)
Open Row to Open Column (typical)
Open Column to Open Column (typical)
Module to Module (Represents 8 x 64 Configuration)*
Closed Path to Closed Path (typical)
Open Row to Open Row (typical)
Open Row to Open Column (typical)
Open Column to Open Column (typical)
*Chaining Cable (part number E1466-80002) used to connect modules
82 Matrix Modules Specifications
W
. AC specifications apply with no more than one
<10 kHz
<100 kHz
- 78 dB
- 57 dB
- 93 dB
- 73 dB
- 84 dB
- 63 dB
- 86 dB
- 65 dB
<10 kHz
<100 kHz
- 78 dB
- 55 dB
- 84 dB
- 66 dB
- 84 dB
- 63 dB
- 93 dB
- 72 dB
W
. AC specifications apply with no more than one
<10 kHz
<100 kHz
- 72 dB
- 50 dB
- 73 dB
- 52 dB
- 84 dB
- 64 dB
- 92 dB
- 70 dB
<10 kHz
<100 kHz
- 66 dB
- 45 dB
- 68 dB
- 46 dB
- 84 dB
- 64 dB
- 92 dB
- 71 dB
W
. AC specifications apply with no more than one
<10 kHz
<100 kHz
- 75 dB
- 54 dB
- 91 dB
- 59 dB
- 85 dB
- 64 dB
- 92 dB
- 71 dB
<10 kHz
<100 kHz
- 72 dB
- 51 dB
- 74 dB
- 53 dB
- 92 dB
- 72 dB
- 82 dB
- 64 dB
<1 MHz
- 41 dB
- 56 dB
- 47 dB
- 48 dB
<1 MHz
- 43 dB
- 52 dB
- 48 dB
- 48 dB
<1 MHz
- 34 dB
- 37 dB
- 47 dB
- 52 dB
<1 MHz
- 29 dB
- 29 dB
- 48 dB
- 52 dB
<1 MHz
- 38 dB
- 43 dB
- 47 dB
- 54 dB
<1 MHz
- 33 dB
- 38 dB
- 56 dB
- 50 dB