Affi Xing The Tefl On Film - YOKOGAWA DPharp EJA110A User Manual

Eja series differential pressure and pressure transmitters
Hide thumbs Also See for DPharp EJA110A:
Table of Contents

Advertisement

IMPORTANT
The transmitter should be installed below the high
pressure (HP) process connection to ensure a
positive head pressure of fi ll fl uid. The recommended
height between the HP process connection and the
transmitter is show in the following table. Pay special
attention to vacuum applications.
Model
EJA118
EJA438W Wetted parts material code S,
T, U, EJA438N
EJA438W Wetted parts material code H
Wetted parts material code
If it can not be installed below the HP process
connection with the recommended height above,
please use the equation below:
(P–P0)×dHg
h=
×7.5×10
ds
h: Vertical height between the HP process connection
and the transmitter (mm)
h≤0: Install the transmitter at least h (mm) below
the HP process connection
h>0: Install the transmitter at most h (mm) above
the HP process connection
P: Pressure in the tank (Pa abs)
P0: Minimum working pressure limit of the transmitter
(ambient temperature range: -10 to 50°C)
EJA118
Wetted parts material
S
code
P0: Minimum working
3178
pressure limit
EJA438
Wetted parts material
S
code
P0: Minimum working
5254
pressure limit
ds: Specifi c gravity of fi ll fl uid (at 25°C)
Fill fl uid code
A
1.07
ds: Specifi c gravity
dHg: Specifi c gravity of the Mercury 13.6 (at 25°C)
Wetted parts material code
Height (h)
600 mm
700 mm
1,300 mm
–3
[mm]
T
H
U
3596
6074
4711
T, U
H
6980
13019
B
C, E
D
1.90 to
0.94
1.09
1.92
Fill fluid code
<3. Installation>
Low
pressure side
h
(+)
0
(–)
Figure 3.8
Example of Installation to Tank
(Caution on Installation)
3.4
Affi xing the Tefl on Film
IMPORTANT
The FEP Teflon option includes a teflon film and
fluorinated oil. Before mounting the transmitter to the
process fl ange, affi x the tefl on fi lm as follows:
1) Position the diaphragm so that the diaphragm is in
an upward position.
2) Pour the fl uorinated oil on the diaphragm and
gasket area covering it completely and evenly. Be
careful not to scratch the diaphragm or change the
its shape.
3) Affi x the tefl on fi lm over the diaphragm and gasket
area.
4) Next, carefully inspect the cover and try to identify
any entrapped air between the diaphragm and
the tefl on fi lm. The air must be removed to ensure
accuracy. If air pockets are present, use your
fi ngers to remove the air by starting at the center of
the diaphragm and work your way out.
5) Place the gasket with the tefl on fi lm and affi x to the
process fl ange.
Fluorinated oil
[PART No. : F9145YN]
Figure 3.9
Affi xing the Tefl on Film
17
P
High
pressure side
F0308.ai
Teflon film
PART No.
EJA210A F9347YD
2B
EJA118W
(50A)
F9347YA
EJA438W
3B(80A)
F9347XA
Diaphragm
Gasket area
F0309.ai
IM 01C22A01-01E
3