3.5 Affixing The Teflon Film - YOKOGAWA EJX110A Installation Manual

Ejx series and eja-e series differential pressure and pressure transmitters
Hide thumbs Also See for EJX110A:
Table of Contents

Advertisement

3.4.2 Mounting to Process Flange
Tighten the bolts to completely close the gap between the
ring and the pressure detector section.
The mating flange, gasket, stud bolts and nuts are to
procured by the customer.
Mating flange
Gasket
Figure 3.10
Mounting to Process Flange
IMPORTANT
• Confirm that there is no gap between the ring
and the process-detector section after they are
mounted on the process flange. A gap can lead to
a sudden, explosive release of process fluids.
• When mounting or removing the ring, take care not
to tilt the pressure detector downward as the ring
can slip off and cause injury.
• When re-mounting the ring, use the new spiral
gasket as shown in below table.
Table 3.1
Spiral Gasket for Pressure Detector
Section Side*
Part number
Size
F9350SV
ø100×ø120×t4.5 For 3-inch flange
F9970XF
ø100×ø120×t4.5 For 3-inch flange**
F9350ST
ø70×ø90×t4.5
F9970XD
ø70×ø90×t4.5
F9346ZH
ø60×ø75×t4.5
F9970XB
ø60×ø75×t4.5
*:
Material; 316SST (Hoop), PTFE Teflon (Filler)
**:
For oil-prohibited use (Option code: /K1, /K2, /K5, /K6)
Spiral gasket
Ring
Diaphragm
Pressure-detector
section
F0310.ai
Description
For 2-inch flange
For 2-inch flange**
For 1 1/2-inch flange
For 1 1/2-inch flange**
<3. Installation>
3.5 Affixing the Teflon Film
IMPORTANT
The FEP Teflon option includes a teflon film and
fluorinated oil. Before mounting the transmitter to the
process flange, affix the teflon film as follows:
1) Position the diaphragm seal so that the diaphragm
is in an upward position.
2) Pour the fluorinated oil on the diaphragm and
gasket area covering it completely and evenly. Be
careful not to scratch the diaphragm or change the
its shape.
3) Affix the teflon film over the diaphragm and gasket
area.
4) Next, carefully inspect the cover and try to identify
any entrapped air between the diaphragm and
the teflon film. The air must be removed to ensure
optimum performance. If air pockets are present,
use your fingers to remove the air by starting at the
center of the diaphragm and work your way out.
5) Position the gasket on the Teflon film.
6) Mount the transmitter onto the process flange.
Fluorinated oil
[PART No. : F9145YN]
Figure 3.11 Affixing the Teflon Film
33
Teflon film
PART No.
EJ210 F9347YD
2 inch
EJ118
F9347YA
EJ438
3 inch
F9347XA
Diaphragm
Gasket area
F0311.ai
IM 01C25A01-01E

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents