Figure 27 Jumper Location On Cpu Pwa - Honeywell 2000 Series Installation, Operation And Maintenance Manual

Herculine 2000 series
Hide thumbs Also See for 2000 Series:
Table of Contents

Advertisement

Set Up and Calibration Procedures
Calibration
Calibrate Input
The HercuLine
1. 0 to 20 mA, or 4 to 20 mA
2. 0 to 5 Volts,1 to 5 Volts, or 0 to 10 Volts
The input type is selected through the Input set up group using the local keypad.
Refer to Figure 25 for the wiring connections and follow the procedure in Table 25 to calibrate the input
circuit of the HercuLine
ATTENTION
For an input calibration to be saved, you must complete the procedure. The calibration will not
be saved if you exit without completing the steps of the procedure.
To exit calibation mode, press DISPLAY or SETUP keys.
68
HercuLine™ 2000 Series Actuator - Installation, Operation and Maintenance Manual
Jumper W2

Figure 27 Jumper Location on CPU PWA

®
2001/2002 actuator accepts a variety of signal inputs.
®
2001/2002 actuator.
Revision 6
7/07

Advertisement

Table of Contents
loading

Table of Contents