Control Unit Mel64/65; Replacing Panel Board - Leica WILD M695 Service Manual

Operating microscope
Table of Contents

Advertisement

Service Manual M695
Doc Code 636 718

Control unit MEL64/65

Replacing panel board

Remove 4 screws with washers (1). Remove
knob cover (2) and knobs (3). Open front cover
(4).
Pull out 2 plugs. Unscrew and remove 4 screws
(5). Remove panel board (6).
Assembly:
The maximum permissible gap between knobs
and front cover is 0.5mm.
Adjust the rotation range of the knobs.
Replacing boards
09/93 3691-DMF