Must I enter both "Found As" and "Left As" Values if the values are the same
when entering a calibration record?
Not always. If the calibration record being entered does not use Separate As
Found/As Left Inputs, then once you enter a "Fnd As" value you may tab to the
corresponding "Lft As" cell and hit the "=" key. This will copy the exact "Fnd
As" value into the "Lft As" cell. Note: The field must be blank for this to work.
After making changes to a record and saving it, I realized that I did not need to
make the changes. Why didn't my Cancel Changes button work?
The Cancel Changes button will work anytime before the changes are saved.
Once a record has been saved though, its previous data cannot be restored.
What is the "Quick Type" feature?
All fields in the system with an associated drop-down list have a "Quick Type"
feature. This means that you may begin typing the desired entry (as long as it is
already on the drop-down list) and DPCTrack2 will automatically insert it into
the field.
Can I print lists generated from Advanced Queries?
Right-clicking on any grid list that you have generated, for example a list
resulting from an Advanced Query, will allow you to export it to a variety of
different formats like a Microsoft Excel® or Word® document. In addition, you
will also see an option to print the list directly from DPCTrack2.
Why am I not able to make changes to the test point grid in the CALIBRATION
screen?
If you would like to make changes to the test point grid in the CALIBRATION
screen, you can do this by clicking the Modify Specs button. This will open up
the grid and allow you to modify any of the cells present. All numeric cells can
be modified by typing over the existing value; all signal type, stated accuracy or
resolution cells can only be modified by activating the cell's drop-down list. In
addition, clicking the Recalculate button will recalculate the high/low spec and
/or control limits if those are affected by your changes. To make the grid
inaccessible again, click the Protect Specs button and continue to enter the rest of
your data. However, performing these actions will not update the item's master
record. Permanent changes must actually be made in the item's master screen.
Please keep in mind though that several settings such as Prevent Accuracy
Modifications During Calibration in the COMPANY RECORD screen, selection
of a "Linear" or "Square Root" Correlation to Output for an item may affect your
ability to modify various information in the calibration specifications grid.