18.3.2 Set Calibrator Date/Time to Current Date/Time

If supported, this menu item will call a function which resets the internal clock of the
calibrator device to your PC's current system date and time.

18.3.3 View Number of Test Results

If supported, a message will appear indicating the number of calibration results that are
contained within the calibrator's memory.

18.3.4 View/Edit Session Log

DPCTrack2 keeps a useful log of Uploads, Downloads and Memory Clears in a log. See
Session Logging for more information.

18.4 Lists

Many of the standard fields in DPCTrack2 have drop-down Lists that are User Defined.
This gives you the ability to customize many areas of the program to meet your needs.
All of the Standard Lists can be accessed from the Lists menu. For further information
about lists see Chapter 12.

18.5 Help

See Chapter 2 for more on the Help menu.

18.6 Options

During any Download or Upload operation, the Options screen allows default values to
be set for tasks or results created. After modifying options the user may select Save as
Default to save the current options as the default options for the next time the open the
screen. The Reset button will return the current set of options to the previously saved
defaults.

18.6.1 Download Options

Options for transferring items as tasks to the Documenting Calibrator.

18.6.1.1 Default Loop Power

The Fluke calibrators have a power source available for transmitters and other
instruments/loops that may need a separate power source. Choose your desired setting
from the dropdown list, and all downloaded tasks will default to this setting when they
are performed on the calibrator.

18.6.1.2 User Entered Values (UEV)

A user-entered value allows the technician to overwrite the actual measured or sourced
value with another numeric value, which is stored as the result. This value is used as the
value stored in DPCTrack2. Choose either Source or Measure or both Source and
Measure UEV to enable Source, Measure or both user-entered values. User-entered
values can be useful for recording information that is read by some other means than the
calibrator's actual sourcing or measuring capabilities