Download Print this page

Meter Modes - Fluke 190-502/AM Technical Data Manual

Datasheet
Hide thumbs Also See for 190-502/AM:

Advertisement

Meter modes

Meter inputs
Via 4 mm banana inputs, fully isolated from
scope inputs and scope ground
Number of readings
One at a time
Maximum resolution
5,000 counts
Input impedance
1 MΩ ± 1 % // 14 pF ± 2 pF
Advanced meter functions
Auto/manual ranging, relative measurements (Zero reference), TrendPlot
The specified accuracy is valid over the temperature range 18 °C to 28 °C
Add 10 % of specified accuracy for each degree C below 18 °C or above 28 °C
Voltage
V dc accuracy
± (0.5 % + 5 counts)
V ac true rms accuracy
± (1 % + 10 counts)
15 Hz to 60 Hz:
± (2.5 % + 15 counts)
60 Hz to 1 kHz:
60 Hz to 20 kHz:
V ac+dc true rms accuracy
± (1 % + 10 counts)
15 Hz to 60 Hz:
± (2.5 % + 15 counts)
60 Hz to 1 kHz:
60 Hz to 20 kHz:
Voltmeter ranges
500 mV, 5 V, 50 V, 500 V, 1,000 V
Resistance
Ranges
500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ
Accuracy
± (0.6 % + 5 counts)
Other meter functions
Continuity
Beeper on < 50 Ω (± 30 Ω)
Diode test
Up to 2.8 V
Current (A)
A dc, A ac, A ac+dc using an optional current clamp or shunt
Scaling factors: 0.1 mV/A, 1 mV/A to 100 V/A and 400 mV/A
Temperature
With optional accessories. Scale factors 1 °C/mV or 1 °F/mV
4 Fluke Corporation
ScopeMeter 190 Series II
Shop for Fluke products online at:
190-062
190-102
MyFlukeStore
www.
190-202
190-502
.com
190-104
190-204
Via BNC scope inputs
Up to 4 simultaneously
999 counts
recording
± (1.5 % + 5 counts)
± (1.5 % + 10 counts)
± (2.5 % + 15 counts)
± (1.5 % + 10 counts)
± (2.5 % + 15 counts)
1.877.766.5412

Advertisement

loading