Specifications And Environmental Conditions; Specifications; Environmental Conditions; Warranty - Fluke 9112A User Manual

Calibration furnace
Table of Contents

Advertisement

3
Specifications and Environmental
Conditions
3.1

Specifications

Operating Range
Stability
Uniformity
Stabilization Time
Heater Power
Power Requirements
Outside Dimensions
Weight
3.2

Environmental Conditions

Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
• temperature range: 5 - 40°C (41 - 104°F)
• ambient relative humidity: 15 - 50%
• pressure: 75kPa - 106kPa
• mains voltage within ± 10% of nominal
• vibrations in the calibration environment should be minimized
• altitude less than 2000 meters
• indoor use only
3.3

Warranty

Fluke Corporation, Hart Scientific Division (Hart) warrants this product to be
free from defects in material and workmanship under normal use and service

3 Specifications and Environmental Conditions

300°C to 1100°C
300°C
±0.05°C
500°C
±0.05°C
700°C
±0.1°C
1000°C
±0.1°C
1100°C
±0.1°C
300°C
±0.05°C
500°C
±0.08°C
700°C
±0.2°C
1000°C
±0.25°C
1100°C
±0.3°C
Typically 2 hours midrange, slower at the low temperature end (4
hours), faster at the high temperature end
3700 Watts High
230 VAC (±10%), 50/60 Hz, 20 A
18" H x 14.25"W x 26"D (457mm x 362mm x 660mm)
72.5 lbs
Specifications
11