Specifications And Environmental Conditions; Specifications; Environmental Conditions - Fluke Hart Scientific 9150 User Manual

Portable furnace
Hide thumbs Also See for Hart Scientific 9150:
Table of Contents

Advertisement

3
Specifications and Environmental
Conditions
3.1

Specifications

Temperature Range
Display Resolution
Stability
Display Accuracy
Well Diameter
Well Depth
Heating Time
Cooling Time
Well to Well Uniformity
Stabilization
Power
Size
Weight
Fault Protection
3.2

Environmental Conditions

Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
• temperature range: 5 - 50°C (41 - 122°F)
• ambient relative humidity: 15 - 50%
• pressure: 75kPa - 106kPa
• mains voltage within ± 10% of nominal
• vibrations in the calibration environment should be minimized
• altitude less than 2000 meters

3 Specifications and Environmental Conditions

150–1200°C (302–2192°F)
0.1° to 999.9°, 1° above 1000°
±0.5°C
±5.0°C
1.25" (32 mm)
4" (102 mm)
35 minutes to 1200°C
140 minutes with block
±0.5°C to ±1.0°C (insert "C" at 1200°C)
20 minutes
115 VAC (±10%), 10.5 amps [230 VAC (±10%), 5.2 amps] switchable,
50/60 Hz, 1200 W
12.4" H x 8.2" W x 12.4"D (315 x 208 x 315 mm)
28 lb. (13 kg)
Sensor burnout and short protection, over temperature thermal cutout
Specifications
9