1
Solution
Observed
4
—
20.2
A
27.7
B
33.0
C
44.5
Method Comparison 1
40
30
20
10
0
0
10
Kit A, µmol/L
(IML 2000) = 1.0 (Kit A) + 0.024 µmol/L
r = 0.966
Method Comparison 2
50
40
30
20
10
0
0
10
HPLC, µmol/L
(IML 2000) = 0.97 (HPLC) + 0.71 µmol/L
r = 0.974
IMMULITE 2000 Homocysteine (PIL2KHO-14, 2007-11-20)
2
3
Expected
%O/E
—
—
25.4
109%
31.7
104%
44.2
101%
20
30
40
20
30
40
Method Comparison 3
4
20
15
10
5
(IML 2000) = 0.90 (Kit B) – 0.02 µmol/L
r = 0.925
Effect of Storage Temperature on
Different Sample Types
Graph 1: Heparin
Graph 2: EDTA
50
5
10
15
Kit B, µmol/L
20
25
7