6 Push the switch into place, with the pins located under the rocker switch. Ensure the switch
clicks securely into place.
Switch - VPAP III, III ST
7 Take the humidifier connector and slide it into the housing, ensuring it "clicks" into place.
8 Locate the wires in the route shown below.
Route cable
behind
second clip
Blue wire
must be
on top
Slot both
wires
between
posts
9 Refit the PCB insulator.
10 Replace the Main PCB and blower.
36
VPAP® III, VPAP III ST, VPAP III ST-A Service Manual
Switch - VPAP III ST-A
VPAP III, VPAP III ST
VPAP III ST-A
Route cable
under clip
Caution
Wires must not
overlap or
!
cross over ribs
Slot brown
wire between
posts on either
side of fuse
casing
Slot both wires
between posts