Installation; Unpacking The Analyzer - Teledyne 3300TA Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

Trace Oxygen Analyzer
Installation of the analyzer includes:
1. Unpacking the system.
2. Mounting the Control Unit to a 19" Rack
3. Installing the Micro-Fuel Cell
4. Making the electrical connections.
5. Making the gas connections.
6. Testing the installation.
CAUTIONS:
3.1
As soon as you receive the instrument, carefully unpack and inspect the
Unit, and any included accessories for damage. Immediately report any
damage to the shipping agent. The analyzer is shipped with all the materials
you need to install and prepare the system for operation.
CAUTION: Do not disturb the integrity of the cell package until the cell is to

Installation

Read this chapter in its entirety before installing the units.
The Sample must be free of entrained solids or water. How-
ever, a high humidity sample is ideal, since it will prevent
water loss from the cell electrolyte.
The Micro-Fuel Cell sensor electrolyte is caustic. Do not
attempt to open it. Leaking or exhausted cells should be
disposed of in accordance with local regulations. Refer to the
Material Safety Data Sheet in the Appendix.
Any damage or scarring of the delicate permeable membrane
on the sensing end of the cell will require cell replacement.
Prevent contact with membrane by any solid object.

Unpacking the Analyzer

actually be used. If the cell package is punctured and air is
permitted to enter, cell-life will be compromised.
Teledyne Analytical Instruments
Installation 3
3-1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents