Rotating Transmitter Section; Changing The Direction Of Integral Indicator - YOKOGAWA DpharpEJX vigilantplant EJX110 User Manual

Differential pressure and pressure transmitters
Table of Contents

Advertisement

4.5 Rotating Transmitter Section

The transmitter section can be rotated approximately
360 (180 to either direction or 360 to one direction
from the original position at shipment, depending on
the configuration of the instrument.) It can be fixed at
any angle within above range.
1) Remove the two setscrews that fasten the transmitter
section and capsule assembly, using the Allen
wrench.
2) Rotate the transmitter section slowly and stop it at
designated position.
3) Tighten the two setscrews to a torque of 1.5 N·m.
IMPORTANT
Do not rotate the transmitter section more than
the above limit.
Vertical impulse piping type
Pressure-detector section
Horizontal impulse piping type
Transmitter section
Figure 4.7
Rotating Transmitter Section (Left Side High
Pressure Type)
Stopper
Rotate 0 to 180 segments
Conduit connection
Transmitter section
Rotate 0 to 180 segments
Conduit connection
Zero-adjustment screw
Pressure-detector section
F0407.EPS
4.6 Changing the Direction of
Integral Indicator
IMPORTANT
Always turn OFF power, release pressure and
remove a transmitter to non-hazardous area
before disassembling and reassmbling an
indicator.
An integral indicator can be installed in the following
three directions. Follow the instructions in section 8.4
for removing and attaching the integral indicator.
Figure 4.8
Integral Indicator Direction
4-4
4 . INSTALLATION
F0408.EPS