Teledyne 311 Instruction Manual page 15

Trace oxygen analyzer
Table of Contents

Advertisement

Procedure II:
1.
Purge the instrument with Nitrogen at the sample port.
2.
Note the reading once it has stabilized (at least 24 hrs. on the 0-10 ppm range) and
increase the flow rate.
3.
If the reading goes down, the unit, or the tubing to the unit, has a leak.
-12-