A&D UA-1010 Instruction Manual

A&D UA-1010 Instruction Manual

Digital blood pressure monitor

Advertisement

Digital Blood Pressure Monitor
Model UA-1010
Instruction Manual
Original
'
Manuel d
instructions
Manual de Instrucciones
Manuale di Istruzioni
使用手冊
1WMPD4002223A

Advertisement

Table of Contents
loading

Summary of Contents for A&D UA-1010

  • Page 1 Digital Blood Pressure Monitor Model UA-1010 Instruction Manual Original ’ Manuel d instructions Manual de Instrucciones Manuale di Istruzioni 使用手冊 1WMPD4002223A...
  • Page 2: Table Of Contents

    Contents Dear Customers..............2 Preliminary Remarks ............. 2 Precautions................2 Parts Identification..............3 Symbols................4 Operation Mode..............6 Using the Monitor ..............7 Installing / Changing the Batteries ........7 Connecting the Air Hose ..........7 Connecting the AC Adapter..........7 Adjusting the Built-in Clock ..........
  • Page 3: Dear Customers

    Dear Customers Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. We recommend that you read through this manual carefully before using the device for the first time.
  • Page 4: Parts Identification

    Parts Identification DC Jack Display Button Button SET Button Air Connector Plug START Button Air Socket Air Hose Proper Fit Range Artery Position Mark Index Mark Arm Cuff Battery Compartment Battery Cover 1.5V Batteries (R6P, LR6 or AA) Display Average MEMORY I.H.B.
  • Page 5: Symbols

    Symbols Symbols that are printed on the device case Symbols Function / Meaning Recommended Action Standby and Turn the device on. Battery installation guide Direct current Serial number Date of manufacture 2010 Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks.
  • Page 6 Symbols Symbols Function / Meaning Recommended Action Previous measurements stored in memory. Average data The battery power indicator during measurement. Full Battery Replace all batteries with new The battery is low when it blinks. ones when the mark blinks. Low Battery Take another measurement.
  • Page 7: Operation Mode

    Operation Mode 1. Normal Measurement Press the START button. Blood pressure is measured and the data is stored in memory. This device can store the last 90 measurements in memory. 2. Recalling the Data Press the button to recall the data in memory.
  • Page 8: Using The Monitor

    Using the Monitor Installing / Changing the Batteries Step 1 1. Remove the battery cover. Step 3 Step 1 2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking Step 2 care that the polarities (+ and -) are correct.
  • Page 9: Adjusting The Built-In Clock

    Using the Monitor Adjusting the Built-in Clock Press Adjust the clock prior to use. 1. Press the SET button until the year starts Year blinking. 2. Select the year using the button. Press the SET button to set the current year and move to month/day selection.
  • Page 10: Selecting The Correct Cuff Size

    Using the Monitor Selecting the Correct Cuff Size Using the correct cuff size is important for an accurate reading. If the cuff is not the proper size, the reading may yield an incorrect blood pressure value. The arm size is printed on each cuff. The index and proper fit range, on the cuff, tell you if you are applying the correct cuff.
  • Page 11: How To Take Proper Measurements

    Using the Monitor Symbols that are printed on the cuff Symbols Function/Meaning Recommended Action Set the mark on the artery of the upper arm or in line with Artery Position Mark the ring finger on the inside of the arm. ▲...
  • Page 12: Measurements

    Measurements Before measurement, read “Notes for Proper Measurement” on the next page. Normal Measurement Place the cuff on the arm (preferably the left arm). At heart level Sit quietly during measurement. Press the button. START All of the display segments are displayed. Press START Zero (0) is displayed blinking briefly.
  • Page 13: Measurement With The Desired Systolic Pressure

    Measurements Measurement with the Desired Systolic Pressure Model UA-1010 is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically. Use this method when re-inflation occurs repeatedly or when the results are not displayed even if the pressure decreases to 20 mmHg or less.
  • Page 14: Recalling The Memory Data

    Recalling the Memory Data Note: This device stores the last 90 measurements in memory. Press 1. Press the button. The average of all measurements and the Average systolic number of data are displayed. (If no data, “0” is displayed. Press the Average diastolic or START button to turn the Average pulse...
  • Page 15: What Is An Irregular Heartbeat

    What is an Irregular Heartbeat The UA-1010 blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you are relaxed, remain still and do not talk during measurements.
  • Page 16: About Blood Pressure

    About Blood Pressure What is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).
  • Page 17: Troubleshooting

    described in this manual to get to know your normal blood pressure. Regular readings give a more comprehensive blood pressure history. Be sure to note the date and time when recording your blood pressure. Consult your doctor to interpret your blood pressure data. Troubleshooting Problem Possible Reason...
  • Page 18: Maintenance

    However it is generally recommended to have the device inspected every 2 years, to ensure proper functioning and accuracy. Please contact the authorized dealer in your area or A&D for maintenance. Technical Data Type UA-1010 Measurement method Oscillometric measurement Measurement range Pressure: 0 - 299 mmHg...
  • Page 19 A&D Company, Limited 1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LTD. Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D Australasia Pty Ltd.