Device Errors 
Error or symbol 
Device shows "EE"  
Device shows  
"E 
  E" vertically in two lines 
Device shows 'E' 
Device shows "Battery 
picture with a cross mark" 
No Power 
 
No Display appears on the 
unit 
Cause 
Cuff is under‐inflated, or has 
not been applied correctly. 
Cuff is over‐inflated.  
Monitor could not detect 
pulse wave. 
Air plug is not connected 
Batteries are worn 
Power off 
Resolution 
Single Mode: 
The monitor will attempt 1 reading. 
 
TrueRead™ Mode: 
The monitor will attempt up to 5 times  
 
After  the automatic attempts, remove 
the arm cuff.  Read, "Taking a 
Measurement" in the instruction manual.  
Take another measurement. 
Check air plug connection.  Take another 
measurement and remain still until the 
measurement is complete. 
Replace the four batteries. 
Refer to "Battery Installation" in the 
instructionj manual. 
Check that the device is turned on.  
 
Check the battery installation for proper 
placement of the battery polarities. 
 
Replace all four batteries with new ones. 
 
If running from a wall outlet, check that 
the cord is attached to both the device 
and outlet 