OCX 8800
Figure 7-7. Alignment of
COe Sensor Assembly
7-10
6. See Figure 7-7. Using straightedge on the sensor holder flat, as shown,
matchmark upper flange of sensor housing to show correct alignment of
sensor holder.
7. With one wrench holding eductor elbow (12, Figure 7-6), and one
wrench on flats of sensor holder (11), unthread and remove COe sensor
assembly (5). Do not allow eductor elbow to turn.
8. Slide band heater (10) and heater insulator (9) from sensor holder (11).
Sensor
Holder
Flat
Matchmark
Matchmark
Straightedge
Instruction Manual
IM-106-880, Rev 1.0
January 2007