Figure 7-24. Solenoid
Power Terminals
7-30
TOP VIEW
#1
#1
Install Cover
1. If removed, install cover gasket (4, Figure 7-20). Screw cover (3) onto
electronics housing. Tighten cover firmly. Align locking clip (2) with gap
between cover ribs.
2. To comply with explosion-proof requirements, cover gasket (4) must be
compressed. For hazardous area applications, verify that the cover
gasket is compressed.
3. Loosen screw (1) and slide locking clip (2) fully into gap between cover
ribs. Retighten screw (1).
Instruction Manual
IM-106-880, Rev 1.0