Specifications; Initial Preparation - Hanna Instruments HI 98701 Instruction Manual

Printing and logging k-j-t thermocouple thermometers
Hide thumbs Also See for HI 98701:
Table of Contents

Advertisement

SPECIFICATIONS

Range (*)
K
-200.0 to 999.9 °C ; 1000 to 1370 °C
-300.0 to 999.9 °F ; 1000 to 2500 °F
J
-200.0 to 760.0 °C
-300.0 to 999.9 °F ; 1000 to 1400 °F
T
-200.0 to 400.0 °C
-300.0 to 750.0 °F
Resolution
K 0.1°C (-99.9 to 999.9 °C); 1°C (1000 to 1370 °C)
0.2 °C (-200.0 to -100.0 °C)
0.2°F (-199.9 to 999.9 °F); 1°F (1000 to 2500°F)
0.3 °F (-300.0 to -200.0 °F)
J 0.1 °C (-149.9 to 760.0 °C)
0.2 °C (-200.0 to -150.0 °C)
0.1 °F (32.0 to 999.9 °F); 1 °F (1000 to 1400 °F)
0.2 °F (-300.0 to 32.0 °F)
T 0.1 °C (-99.9 to 400.0 °C)
0.2 °C (-200.0 to -100.0 °C)
0.1°F (300.0 to 750.0°F);0.2°F (-149.9 to 300.0°F)
0.3 °F (-300.0 to -150.0 °F)
Accuracy
±0.5 °C (-200.0 to 999.9 °C); ±1°C outside
(@20°C/68°F)
±1°F
for one year (excluding probe error)
Typical EMC Dev.
±5 °C; ±9 °F
Channels
HI 98701 / HI 98801: 1 channel
HI 98704 / HI98804: 4 channels
Probe
K , J or T-type thermocouple
Cold Junction
NTC 10K ; 0.1°C resolution ; ±0.3°C accuracy
Printer
Low power impact, 14 characters per line using
38 mm plain paper (HI 710034)
Printing/Logging
Selectable at
Interval
1, 2, 5, 10, 15, 30, 60, 120 or 180 minutes
Auto-off
Selectable after 5, 10, 15, 30, 45 or 60 minutes
Power supply
4 x 1.5V AA batteries / approx. 350 hours of use
(without printing and backlight); or 12 Vdc input
Environment
0 to 50°C (32 to 122°F); RH max 95%
Dimensions
220 x 82 x 66 mm (8.7 x 3.2 x 2.6")
Weight
550 g (1.2 lb.)
(*)
Tables from NIST Monograph 175 revised to ITS-90 are used
6

INITIAL PREPARATION

Each meter is supplied complete with batteries. Remove the back
cover, unwrap the batteries and install them while paying attention
to the polarity.
Alternatively, connect a 12 Vdc voltage adapter to the power adapter
plug on the top of the instrument.
To prepare the instrument for use, choose the most appropriate
temperature probe(s) for your application (see "Accessories" section for
a complete list) and connect it (them) to the connector(s) on the top
of the instrument.
With the meter facing you, channel #1 is the first connector on the
top left hand side.
To switch the meter on, press the ON/OFF key.
The batteries charge status or "LINE" message
(if external power adapter is connected) will be
displayed on the LCD for a few seconds.
The meter is now ready to operate.
To maximize battery life, the meter is automatically switched off after a
user selectable period of non-use (this feature is enabled and set to 5
minutes by default; it can be disabled or changed through setup
code #40).
If in logging mode, after the period of non-use, the meter will
continue to monitor the temperature periodically at the end of every
logging interval. Only the "LOG" indication will be displayed.
While storing data in memory, during the sleep mode, the reading
will appear briefly on the LCD.
To reactivate the display press the ON/OFF key.
Note: When the use of an alternate function (FNC, CFM and LOT
INFO) is requested, press and hold the ALT key first and then
the second key.
ON/OFF
7