Adjusting Ir-Vl Ir-Fusion® Blend Level - Fluke TiR4 User Manual

Ir flexcam thermal imagers
Hide thumbs Also See for TiR4:
Table of Contents

Advertisement

Adjusting IR-VL IR-Fusion® Blend Level
This feature is disabled with optional 10 and 54-mm lenses.
You can change the IR-Fusion® blend level from full infrared (IR) to full
visible light (VL) or some combination in between. When the IR-Fusion®
blend level is set to full IR, 100% of the image is IR. When the IR-Fusion®
blend level is set to full VL, 100% of the image is VL. When the IR-Fusion®
blend level is set somewhere in the middle of full IR and full VL, an
appropriate amount of the image is IR and an appropriate amount is VL, thus
displaying a fused—or blended—image.
To adjust the IR-VL Fusion blend level using the level & span button and
mouse controller:
1. Press and hold G until the IR-Fusion® level dialog box appears near the
bottom of the screen.
2. While continuing to hold G, move the mouse controller left or right to
desired setting, or use E to position the pointer over the Full IR or Full
VL text to quickly move to 100% infrared or 100% visible light
respectively.
3. Tap to accept setting change and return to scan target mode.
To adjust the IR-VL IR-Fusion® blend level from the popup menu:
1. Tap F.
2. Use the mouse controller to position the pointer over Camera Settings on
the popup menu and tap E.
3. Position the pointer over the VLCM tab and tap E.
You may need to position the pointer over the right or left arrow on the tab
bar and tap E to scroll to the VLCM tab.
4. Position the pointer over the Full IR/Full Visible bar; press and hold E
and then move the mouse controller left or right to desired setting.
5. Tap to accept setting change and return to scan target mode.
Visible Light Camera Module (VLCM)
Adjusting IR-VL IR-Fusion® Blend Level
Note
5
5-3

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

Ti45Ti40Ti55Tir3Tir2Ti50

Table of Contents