Model 1587; Model 1577 - Fluke 1577 User Manual

Insulation multimeters
Hide thumbs Also See for 1577:
Table of Contents

Advertisement

Model 1587

Output Voltage
Display Range
0.01 to 6.00 M
50 V
(0 % to + 20 %)
6.0 to 50.0 M
0.01 to 6.00 M
100 V
6.0 to 60.0 M
(0 % to + 20 %)
60 to 100 M
0.1 to 60.0 M
250 V
(0 % to + 20 %)
60 to 250 M
0.1 to 60.0 M
500 V
(0 % to + 20 %)
60 to 500 M
0.1 to 60.0 M
1000 V
60 to 600 M
(0 % to + 20 %)
0.6 to 2.0 G

Model 1577

Output Voltage
Display Range
0.1 to 60.0 M
500 V
(0 % to + 20 %)
60 to 500 M
0.1 to 60.0 M
1000 V
(0 % to + 20 %)
60 to 600 M
Resolution
Test Current
0.01 M
1 mA @ 50 kΩ
0.1 M
0.01 M
0.1 M
1 mA @ 100 kΩ
1 M
0.1 M
1 mA @ 250 kΩ
1 M
0.1 M
1 mA @ 500 kΩ
1 M
0.1 M
1 M
1 mA @ 1 MΩ
100 M
Resolution
Test Current
0.1 M
1 mA @ 500 k
1 M
0.1 M
1 mA @ 1 M
1 M
Insulation Multimeters
Electrical Specifications
Resistance Accuracy
(% of Rdg + Digits)
(3 % + 5 counts)
(3 % + 5 counts)
(1.5 % + 5 counts)
(1.5 % + 5 counts)
(1.5 % + 5 counts)
(10 % + 3 counts)
Resistance Accuracy
(% of Rdg + Digits)
(2.0 % + 5 counts)
(2.0 % + 5 counts)