Gas Flow 0-5 Vdc Analog Output Calibration; Gas Flow 4-20 Ma Analog Output Calibration - Omega FMA 4000 User Manual

Digital mass flow meters
Table of Contents

Advertisement

Enter Backdoor mode by typing:
Unit will respond with:
Disable DAC update by typing:
Unit will respond with:

7.3.2 Gas flow 0-5 Vdc analog output calibration

1.
Install jumpers J7A, J7B and J7C on the PC board for 0-5 Vdc output (see Table VI).
2.
Connect a certified high sensitivity multi meter set for the voltage measurement to the
pins 2 (+) and 1 (-) of the 15 pins D connector.
3.
Write 4000 counts to the DAC channel 1: !11,WRITE,1,4000[CR]
4.
Read voltage with the meter and calculate:
5.
Save FlowOutScaleV in to the EEPROM:
Where: X – the calculated AoutScaleV value.

7.3.3 Gas flow 4-20 mA analog output calibration

1.
Install jumpers J7A, J7B and J7C on the PC board for 4-20 mA output (see Table VI).
2.
Connect a certified high sensitivity multi meter set for the current measurement to
pins 2 (+) and 1 (-) of the 15 pins D connector.
3.
Write 4000 counts to the DAC channel 1:
4.
Read current with the meter and calculate:
5.
Write zero counts to the DAC channel 1:
6.
Read offset current with the meter and calculate:
7.
Save AoutScale_mA in to the EEPROM:
Save AoutOffset_mA in to the EEPROM:
Where: Y – the calculated AoutScale_mA value.
Z – the calculated AoutOffset_mA value.
Note: When done with the analog output calibration make sure the
DAC update is enabled and the BackDoor is closed
(see command below).
!11,MW,1000,1[CR]
!11,BackDoorEnabled: Y
!11,WRITE,4,D[CR]
!11,DisableUpdate: D
!11,MW,25,X[CR]
!11,WRITE,1,4000[CR]
!11,WRITE,1,0CR]
!11,MW,27,Y[CR]
!11,MW,28,Z[CR]