Analog Output Calibration Of Fma 4000 Mass Flow Meter - Omega FMA 4000 User Manual

Digital mass flow meters
Table of Contents

Advertisement

If a simple communication terminal is used for communication with the FMA 4000,
then "MW" (Memory Write) command from the software interface commands set
may be used to adjust sensor value in the linearization table (see section 8.3 for
complete software interface commands list).
Memory Read "MR" command can be used to read the current value of the index.
Assuming the FMA 4000 is configured with RS485 interface and has address
"11", the following example will first read the existing value of Index 133 and then
write a new adjusted value:
!11,MR,133[CR]
!11,MW,133,3450[CR] - writes new sensor value (3450 counts) in to the index 133
Once 100% F.S. calibration is completed, the user can proceed with calibration for
another 9 points of the linearization table by using the same approach.
Note: It is recommended to use Omega
maintenance software for gas table calibration. This software includes
an automated calibration procedure which may radically simplify
reading and writing for the EEPROM linearization table.
7.3
Analog output Calibration of FMA 4000
Mass Flow Meter
FMA 4000 series Mass Flow Meters are equipped with
calibrated 0-5 Vdc and 4-20 mA output signals. The set
of the jumpers (J7A, J7B, J7C) on the printed circuit
board is used to switch between 0-5 Vdc and 4-20 mA
output signals (Figure c-1 "FMA 4000 configuration
jumpers").
FUNCTION
ANALOG
0-5 VDC
OUTPUT
4-20 mA
RS485
TERMINAL
RESISTOR
Figure c-1 FMA 4000 Analog Output Configuration Jumpers
- reads EEPROM address 133
J7A
5-9
1-5
OFF
ON
®
JCD
J7B
J7C
6-10
7-11
2-6
3-7
8-12
4-8
23
supplied calibration and
AutoZero/Reset
push button.
J7 Jumpers