Fluke 105B User Manual page 162

Scopemeter
Table of Contents

Advertisement

Appendixes
Safety
Designed to Installation Category III per IEC 1010,
CSA C22.2 (including approval), and UL1244
(including approval) for 600V measurements on
industrial power distributions systems.
MAX. Input voltage on ................300V rms, 600V rms with
INPUT A or INPUT B
MAX. Input voltage on EXTernal input ................ 300 V rms
Surge Protection ....................... 4 kV, 6 kV with 10:1 probe
per IEC 664 and ANSI/IEEE C62.41 test method
Maximum Isolation to Earth .....600V rms from any terminal
Optical Interface .................. Optically Isolated to 600V rms
ACCESSORY INFORMATION
Passive Probe 10:1 (PM8918)
Attenuation: x10
Useful Bandwidth: 100 MHz
Input resistance: 10 MΩ
Input capacitance: 15 pF
Max. V x Hz: 10
10:1 probe
Cable length: 1.5m
Max. signal Voltage: 600V rms
Max. surge Voltage: 6 kV
Max. Voltage to Ground: 600V rms
Double insulated per IEC1010-1 for 600V.
Optically Isolated RS-232 Adapter /Cable (PM9080/001)
Serial Printer Interface
Full PC Interface
Capture Screen.........................With FlukeView™ software
9
92B
.......................EPSON FX/LQ,
Deskjet Thinkjet, Deskjet, Laserjet
Print screens and waveforms or
log meter and cursor readings
92B 96B
.......................Transfer setups,
waveforms, measurements, etc.
(optional for Fluke 92B, 96B and 99B)
9 - 13