Konica Minolta CA-210 Brochure page 6

Display color analyzer

Advertisement

Dimensions
(Units : mm)
Main Body
340
Specifications
Item
CA-210 (Universal Measuring Probe)
Receptor
Detector: Silicon photo cell
φ27 mm
Measurement area
±2.5˚
Acceptance angle
Pointing function
By LED
Measurement distance
30±10 mm
Display range
Luminance
0.01 to 1000 cd/m
Chromaticity
Displayed in 4 or 3-digit value (Can be chosen)
Luminance
Measurement range
0.10 to 1000 cd/m
Accuracy (for
±2%±1 digit of reading (temperature : 23°C±2°C,relative humidity : (40±10)%)
white)∗1
Repeatability(2 σ)
0.10 to 0.99 cd/m
∗1
1.00 to 1000 cd/m
Chromaticity
Measurement range
0.10 to 1000 cd/m
Accuracy∗1
0.10 to 4.99 cd/m
(temperature : 23°C±2°C,
5.00 to 39.99 cd/m
relative humidity : (40±10)%)
40.00 to 1000 cd/m
160 cd/m
(±0.004 for
Repeatability∗1
0.10 to 0.19 cd/m
0.20 to 0.49 cd/m
0.50 to 1.99 cd/m
2.00 to 1000 cd/m
Flicker Contrast Measurement range
method
Display range
Accuracy
Repeatability
Flicker JEITA
Measurement range
method
∗3
Accuracy
Repeatability
Measurement
xyLv∗4
5 (4.5)
speed
20 (17 ) 4.00 cd/m
(measurements/sec. ) Flicker Contrast
Flicker
JEITA∗3
Display
Digital
xyLv, XYZ, T∆uvLv, u'v'Lv, RGB analyze
Chromaticity is displayed up to fourth decimal place. (Three-digit indication can be chosen.)
Analog
∆x∆y∆Lv, R/G B/G ∆G, ∆R B/R G/R
LCD
16 characters by 2 lines (with backlight)
SYNC mode
NTSC, PAL, EXT, UNIV, INT
Object under measurement
Vertical synchronizing frequency: 40 to 200 Hz
Memory channel
100 channels
Analyzer function
Standard function
Interface
RS-232C (38,400 bps or below), USB (Rev.1.1)
Multi-point Measurement
Max. 5 points(Use 4-Probe Expansion Board CA-B14)
Software
SDK software (supplied as standard accessory)
Operating temperature/humidity range
Temperature : 10 to 28°C; relative humidity 70 % or less with no condensation
Chromaticity change ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta's standard LCD∗1, 160.0 cd/m
Storage temperature/humidity range
0 to 28°C : relative humidity 70 % or less with no condensation 28 to 40°C : relative humidity 40 % or less with no condensation
Input voltage range
100 – 240 V
Main body: 340 (W) × 127 (H) × 216 (D) mm,
Size
Probe: φ49 × 204 mm
Weight
Main body: 3.58 kg, Probe: 520 g
∗1
: The chromaticity and luminance are measured under Konica Minolta's condition (standard LCD(6500 K, 9300 K) is used).
∗2
: The luminance for monochrome is measured when the reading of luminance for white is 160 cd/m
∗3
: Measurement of flicker (JEITA method) is supported by SDK software.
∗4
: Measuring probe connected to probe connector P1 only,used USB (used RS-232C Baud rate: 38400 bps)
∗5
: Measured by Konica Minolta's PC (P3-600 MHz)
SAFETY PRECAUTIONS
For correct use and for your safety, be sure to read
the instruction manual before using the instrument.
Always connect the instrument to the
specified power supply voltage.
Improper connection may cause a fire
or electric shock.
©2003 KONICA MINOLTA SENSING, INC.
Universal Measuring Probe
LCD Flicker Measuring Probe
ISO screw
Tripod socket
ISO screw
CA-210 (Small Universal Measuring Probe)
φ10 mm
±5˚
30±5 mm
2
0.01 to 3000 cd/m
2
0.30 to 3000 cd/m
2
0.2%+1 digit
0.30 to 2.99 cd/m
2
0.1%+1 digit
3.00 to 3000 cd/m
2
0.30 to 3000 cd/m
2
±0.008 for white
0.30 to 14.99 cd/m
2
±0.005 for white
15.00 to 119.99 cd/m
±0.003 for white
2
120.00 to 3000 cd/m
±0.002 for white
2
160 cd/m
2
monochrome)∗2
(±0.004 for
monochrome)∗2
2
0.015 (2 σ)
0.30 to 0.59 cd/m
2
0.008 (2 σ)
0.60 to 1.49 cd/m
0.003 (2 σ)
2
1.50 to 5.99 cd/m
0.001 (2 σ)
2
6.00 to 3000 cd/m
-------
-------
-------
-------
-------
-------
-------
0.10 to 3.99 cd/m
2
5 (4.5)
0.30 to 11.99 cd/m
2
or higher
20 (17 ) 12.00 cd/m
-------
-------
~
, 50–60 Hz, 50 VA
Main body: 340 (W) × 127 (H) × 216 (D) mm,
Probe: φ49 × 232 mm
Main body: 3.58 kg, Probe: 540 g
φ49
φ30
10
30.2
φ44
CA-210 (LCD Flicker Measuring Probe)
φ27 mm
±2.5˚
30±10 mm
2
0.01 to 1000 cd/m
2
2
0.10 to 1000 cd/m
2
2
2
0.2%+1 digit
0.10 to 0.99 cd/m
0.2%+1 digit
2
0.1%+1 digit
1.00 to 1000 cd/m
2
0.1%+1 digit
2
0.10 to 1000 cd/m
2
2
±0.008 for white
0.10 to 4.99 cd/m
2
±0.005 for white
2
±0.005 for white
2
±0.004 for white
5.00 to 19.99 cd/m
±0.003 for white
±0.003 for white
2
20.00 to 1000 cd/m
2
±0.002 for white
±0.002 for white
160 cd/m
2
(±0.004 for
monochrome)∗2
2
0.015 (2 σ)
0.10 to 0.19 cd/m
2
0.010 (2 σ)
2
0.008 (2 σ)
2
0.005 (2 σ)
0.20 to 0.49 cd/m
0.003 (2 σ)
0.002 (2 σ)
2
0.50 to 0.99 cd/m
2
0.001 (2 σ)
0.001 (2 σ)
2
1.00 to 1000 cd/m
2
5 cd/m
2
or higher
0.0 to 100 %
±1 % (Flicker frequency: 30 Hz AC/DC 10 % sine wave)
±2 % (Flicker frequency: 60 Hz AC/DC 10 % sine wave)
1 % (2 σ) (Flicker frequency: 20 to 65 Hz AC/DC 10 % sine wave)
5 cd/m
2
or higher
±0.5 dB
(Flicker frequency: 30 Hz AC/DC 10 % sine wave)
0.3 dB (2 σ)
(Flicker frequency: 30 Hz AC/DC 10 % sine wave)
2
5 (4.5 )
0.10 to 1.99 cd/m
2
or higher
20 (17 ) 2.00 cd/m
2
or higher
16 measurements/sec. (16 measurements/sec.)
0.5measurements/sec. (0.3 measurements/sec.)
xyLv, XYZ, T∆uvLv, u'v'Lv, RGB analyze
Chromaticity is displayed up to fourth decimal place. (Three-digit indication can be chosen.)
Flicker (Contrast method)
∆x∆y∆Lv, R/G B/G ∆G, ∆R B/R G/R, Flicker (Contrast method)
Vertical synchronization frequency: 40 to 200 Hz (Luminance or chromaticity measurement), 40 to 130 Hz (Flicker measurement)
Luminance change : ±2 % ±1 digit of reading for white
Main body: 340 (W) × 127 (H) × 216 (D) mm,
Probe: φ 49 × 204 mm
Main body: 3.58 kg, Probe: 520 g
.
2
Certificate No : YKA 0937154
Certificate No : JQA-E-80027
Registration Date : March 3, 1995
Registration Date : March 12, 1997
9242-4885-15
Small Universal Measuring Probe
Small LCD Flicker Measuring Probe
φ49
φ26
ISO screw
Tripod socket
10
ISO screw
30.2
φ44
CA-210 (Small LCD Flicker Measuring Probe)
φ10 mm
±5˚
30±5 mm
0.01 to 3000 cd/m
2
0.30 to 3000 cd/m
2
2
0.30 to 2.99 cd/m
0.2%+1 digit
3.00 to 3000 cd/m
2
0.1%+1 digit
0.30 to 3000 cd/m
2
0.30 to 14.99 cd/m
2
±0.005 for white
2
±0.004 for white
15.00 to 59.99 cd/m
±0.003 for white
60.00 to 3000 cd/m
2
±0.002 for white
160 cd/m
2
(±0.004 for
monochrome)∗2
0.30 to 0.59 cd/m
2
0.010 (2 σ)
2
0.005 (2 σ)
0.60 to 1.49 cd/m
0.002 (2 σ)
1.50 to 2.99 cd/m
2
0.001 (2 σ)
3.00 to 3000 cd/m
2
15 cd/m
2
or higher
15 cd/m
2
or higher
2
5 (4.5)
0.30 to 5.99 cd/m
2
20 (17) 6.00 cd/m
2
or higher
∗5
∗3
∗3
2
, with 23°C 40 %
Main body: 340 (W) × 127 (H) × 216 (D) mm,
Probe: φ49 × 232 mm
Main body: 3.58 kg, Probe: 540 g