2.
SPECIFICATIONS, APPROVALS, AND
COMPLIANCE
Ranges
(Physical Analog Output)
Measurement Units
Noise at Zero
Noise at Span
Lower Detectable Limit
Zero Drift
Span Drift
Lag Time
Rise / Fall Time
Linearity
Precision
Sample Flow Rate
AC Power, Analyzer
AC Power, Ext Pump
Analog Output Ranges
Recorder Offset
Standard I/O
Optional I/O
Analog Output Resolution
Temperature Range
Humidity
Hydrocarbon Interference
Voltage Coefficient
Dimensions H x W x D
Weight, Analyzer
07450B DCN6495
Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
PARAMETER
1
1
1
2
2
1
1, 3
SPECIFICATION
Min: 0-5 ppb Full scale
Max 0-2000 ppb Full scale (selectable, independent NO, NO
ranges and auto ranging supported)
3
ppb, ug/m
(user selectable Single, independent or auto ranges)
<25 ppt (RMS)
<0.5% of reading (RMS ) above 5 ppb
50 ppt NO, <100 ppt NO
(RMS)
2
<0.1 ppb / 24 hours
<0.5% of reading or 50 ppt RMS whichever is greater /24 hrs
20 seconds
<50 sec to 95%
1% of full scale or 0.1 ppb whichever is greater
0.5% of reading above 5 ppb
3
1100 cm
/min. ± 10%
100V-120V, 60Hz (165W); 220V-240V, 50 Hz (165W)
115V, 60 Hz (345W); 220V-240V, 50/60 Hz, (230W)
All outputs: 10V, 5V, 1V, 0.1V (selectable)
All ranges with 5% under/over range
± 10%
1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital status outputs (7 defined, 1 spare)
6 opto-isolated digital control inputs (4 defined, 2 spare)
4 analog outputs
1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs
1 part in 4096 of selected full-scale voltage
o
5 - 40
C
o
(20-30
C within drift and noise specifications, and operating with EPA
equivalency)
0-95% RH non-condensing
Propylene rejection ratio > 20,000:1
Ethylene rejection ratio > 40,000:1
< 0.1% per V
7" x 17" x 23.5" (178mm x 432 mm x 597 mm)
40 lbs (18 kg)
, NO
2
X
9