TROUBLESHOOTING GUIDE
Under certain conditions. performance of this unit may be adversely affected by factors other than defects. If problems
occur,
please review the following list and take remedial action as appropriate. If you cannot
solve
the problem
after
checking the entire list, please contact your local Olympus representa tive for assistance.
Problem
I
Cause
Rem
edy
I
Page
I
.
Optical
System
a) Lamp does not
light.
Bulb burned out.
Replace
bulb.
3
Fuse burned out.
Replace fuse.
6
b)
Lamp
lights, but field of
Field
iris
diaphragm
is
not open
Wide
Open the field
IriS
diaphragm.
18
view remains dark.
enough.
Condenser is too low.
Adjust
the condenser position.
18
Light path selector knob is set to the
dtI position.
Move the knob to
thec(~
or
(I{
position.
17
c)
Field of view
is
ob-
Light path selector knob
is
in the middle
Set the knob according to the obselVation
17
scured. or field of view
position.
method.
;s
not evenly
illumi-
nated.
The revolving nosepiece
is
not correctly
engaged.
Make sure that the objective dicks proper-
Iy
into
place.
9
The revolving nosepiece
is
not properly
Press it in as far as
it
will go.
10
mounted.
You are using an objective that falls
Use
,
condenser that matches the
19
outside 01 the condenser's illumination
objective.
range.
The condenser is not property centered.
Center the condenser.
18
The
field
iris
diaphragm
is
stopped down
too far.
Open the field ins diaphragm.
18
Th. halogen bulb ;s not mounted
Push the pins of the halogen bulb
into
3
properly.
the holes.
d)
Dirt or
dust
is visible in
Dirt on the light exit glasss
Clean thoroughly.
the field of view.
Dirt on the top surface of the condenser
-
Dirt/dust on
the
specimen
Dirt/dust on the eyepiece
e) The image
is
too con-
The condenser is
lowered
too
far.
Adjust the condenser
position.
18
trasty and dark.
The aperture iris diaphragm IS stopped
down too far.
Open the aperture Iris diaphragm.
19
f) Visibility is poor.
You are using an objective that is not
Use an objective made for the UIS series.
-
Image IS not sharp.
made for the UIS series.
Contrast
is poor.
Details
are
indistinct.
The revolving nosepiece is not position-
Push the nosepiece slider firmly until it
10
ed properly.
comes up against the slap.
The objective is not property POSitioned
in the light path.
Make sure that the object clicks properly
into
place.
9
The correction collar on the correction
While focusing, turn the correction collar
-
collar
equipped objective
is
not adjusted.
to find the best position.
The tip of the objective IS dirty.
Clean the
objective.
-
Immersion oil is .not being used with
Use oiL
21
an oil immersion lens.
The immersion oil contains bubbles.
Remove the bubbles.
21
25
--------------------------.