Radiographic Working Modes

You can select following radiographic working modes according to the
parameters to be controlled and the degree of automation:
• One Point Mode (1P), by selecting kVp with AEC operations.
• Two Point Mode (2P), by selecting kVp and mAs.
• Three Point Mode (3P), by selecting kV, mA and exposure time
independently.
Topics:
One Point Mode (1P)
Two Point Mode (2P)
Three Point Mode (3P)
One Point Mode (1P)
By selecting one of the AEC field buttons, the one point mode is activated.
The value of kVp, mAs, mA, max ms, the setting of focal spot, density,
sensitivity, patient size and the selected AEC fields can be adjusted.
The value of mAs is the max mAs. The default value for max mAs is the highest
supported value.
The value for ms is not available.
Disabling all AEC fields will switch to two point mode.
After exposure all values reflect the settings actually used by the generator.
Figure 31: 1P working mode
DX-D 300 | Operating the DX-D Software Console | 93
0172G EN 20230627 1651