Start up transmitter

Procedure
1. Wire sensor(s) to the signal boards.
See
Start Guide for additional details. Make current output, alarm
relay, and power connections.
2. Once connections are secured and verified, apply DC power to
the transmitter.
When the transmitter is powered up for the first time, Quick
Start screens will appear. Quick Start operating tips are as
follows:
3. Choose the desired language and press ENTER.
4. Choose measurement and press ENTER.
5. For pH, choose preamplifier location.
6. If applicable, choose units of measurement.
7. For contacting and toroidal conductivity, choose the sensor
type and enter the numeric cell constant using the keys.
Quick Start Guide
Wiring
for wiring instructions. Refer to the sensor Quick
This symbol identifies a risk of electical shock.
This symbol identifies a potential hazard. When this
symbol appears, consult the manual for appropriate
action.
A highlighted field shows the position of the cursor.
To move the cursor left or right, use the keys to the left or
right of the ENTER key. To scroll up or down or to increase
or decrease the value of a digit, use the keys above and
below the ENTER key. Use the Left or Right keys to move
the decimal point.
Press ENTER to store a setting. Press EXIT to leave without
storing changes. Press EXIT during Quick Start to return
the display to the initial startup screen (Select language).
a) Select Analyzer to use the integral preamplifier in the
transmitter.
b) Select Sensor/J-Box if your sensor is SMART or has
an integral preamplifier or if you are using a remote
preamplifier located in a junction box.
Quick Start Guide
33