12.2 Information on electromagnetic compatibility (EMC)
Electromagnetic interference immunity
The RadiForce LS580W was tested with the following compliance levels in accordance with the test
requirements for professional healthcare facilities, as established in IEC/EN 60601-1-2.
Customers and users of the monitor have to ensure that the monitor is used in such an environment.
Interference
Measure-
immunity test
ment level
Line-based dis-
3 V
rms
turbances
150 kHz to
caused by RF
80 MHz
fields
6 V
IEC/EN
rms
ISM bands
61000-4-6
between
150 kHz and
80 MHz
Electromagnetic
3 V/m
RF fields
80 MHz to
2.7 GHz
IEC/EN
61000-4-3
Note: The higher frequency range applies at 80 MHz and 800 MHz.
Note: Guidelines with respect to line-based interference due to RF fields or electromagnetic RF
fields may not apply in all situations. Absorption and reflection by structures, objects, and people im-
pact the propagation of electromagnetic waves. .
a)
The field strengths of fixed transmitters, for example the base station for cordless and mobile tele-
phones, radio, land mobile radio, ham radio, and television cannot be determined precisely in ad-
vance. To evaluate the electromagnetic environment using fixed transmitters, an electromagnetic lo-
cation measurement should be included. If the measured field strength in the environment where the
device is used exceeds the applicable RF compliance level, observe the monitor to ensure its proper
operation. If improper operation is observed, in some circumstances additional measures may be
necessary, such as reorienting or repositioning the device.
38
Compliance
Information regarding the electromagnetic envi-
level
ronment
3 V
Portable and mobile RF communications devices
rms
may only be operated in the vicinity of the monitor
and its components (including cables) when in com-
pliance with the recommended minimum distance. It
6 V
rms
is determined using the formula for calculating the
frequency of the transmitter.
Recommended minimum distance
d = 3.5/3 √P = 1.2 √P, 150 kHz to 80 MHz
3 V/m
d = 2 √P, ISM bands between 150 kHz and 80 MHz
d = 3.5/3 √P = 1.2 √P, 80 MHz to 800 MHz
d = 7/3 √P = 2.3 √P, 800 MHz to 2.7 GHz
In this case, "P" stands for the measured maximum
nominal output power in watts (W) of the transmitter
recommended by the transmitter manufacturer, and
"d" for the recommended minimum distance in me-
ters (m).
The field strengths of fixed transmitters according to
electromagnetic location measurement
less than the compliance level in each individual fre-
quency range.
Interference can occur when used in the vicinity of
devices identified with the following symbol.
a)
have to be
RadiForce LS580W
Instructions for Use, 08/2022

Advertisement

Table of Contents
loading

Table of Contents