Technical Information

Cleaning FRG-700

Cleaning the measuring chamber and the polarity insert
1
Using a polishing cloth rub the inside walls of the measuring chamber and the
polarity insert to a bright finish.
The sealing surfaces must only be worked concentrically.
NOTE
2
Rinse the measuring chamber and the polarity insert with cleaning alcohol.
3
Allow both to dry.
Cleaning or replacing the anode
1
Remove the used ignition aid (10a) with tweezers.
2
Using a polishing cloth rub the anode pin to a bright finish
Do not bend the anode. Do not carry out mechanical work on the ceramic part.
NOTE
3
Rinse the anode with cleaning alcohol.
4
Allow the anode to dry.
5
Insert a new ignition aid (10a) into the mounting tool.
6
Carefully press the anode (clean or new) centered and parallel to the tool axis
into the ignition aid and insert it to a depth of approx. 15 mm. The final
positioning is established after the anode is installed.
Cleaning the Pirani element
1
Remove the FPM seal (13a) from the Pirani element (13).
2
Fill the Pirani measuring tube with cleaning alcohol and let it work.
3
Pour the alcohol out of the tube.
4
Dry the tube (e.g. with a blow dryer <150 °C).
5
Slide a new FPM seal over the Pirani element and insert it into the
corresponding groove.
6
Reinstall the Pirani element.
Agilent Inverted Magnetron Pirani Gauge FRG-700 and FRG-702 User Manual