4 . Select an age with the Age Precision™ Technology button
toggle through each age group .
NOTE: You must select an age in order to take a measurement . Age selection will default to last
age setting if you do not change .
0-3 Months
3-36 Months
The thermometer is ready to take a temperature when the display looks like the image below .
5 . Fully insert the probe into the ear canal and point
straight towards the opposite ear or temple .
Once positioned correctly, press and release the
start button
.
6 . ExacTemp® Stability Indicator light will pulse while temperature measurement is in progress .
The light will remain solid for 3 seconds to indicate that a successful temperature reading has
been achieved .
NOTE: If the probe has been properly inserted into the ear canal during the measurement, a
long beep will sound to signal the completed measurement .
If the probe has not been constantly placed in a stable position in the ear canal, a sequence
of short beeps will sound, the ExacTemp® Stability Indicator light will go out and the display
will show an error message (ERROR + position error animated icon = position error) .
See «Errors and troubleshooting» section for more information.
7 . The confirmation beep indicates that an accurate temperature measurement has been
taken . The result is shown on the display .
0-3
3-36
36+
0-3
3-36
36+
0-3
3-36
36+
Age range
0*-3 Months
3-36 Months
36 Months -
adult
* 0 months is defined as having a gestational age of greater than or equal to 37 calendar weeks at birth .
6
36 Months-
adult
Green
Yellow
Normal temperature
Elevated temperature
≥96 .4 – ≤99 .4 °F
(≥35 .8 – ≤37 .4 °C)
≥95 .7 – ≤99 .6 °F
>99 .6 – ≤101 .3 °F
(≥35 .4 – ≤37 .6 °C
(>37 .6 – ≤38 .5 °C)
≥95 .7 – ≤99 .9 °F
>99 .9 – ≤103 .0 °F
(≥35 .4 – ≤37 .7 °C)
(>37 .7 – ≤39 .4 °C)
as shown . Press button to
Red
High temperature
>99 .4 °F
(>37 .4 °C)
>101 .3 °F
(>38 .5 °C)
>103 .0 °F
(>39 .4 °C)