5
Miscellaneous
5.1

Checks

PMF
Cassette locking
Check that the camera used is properly locked to the object tables.
PMA
UI's and SPEED Infos
Check whether all relevant UI's and SPEED Infos have been completed.
Check that the customer performs backup regularly.
5.2

Final tests

SIE
Protective conductor test
Perform the protective conductor test according to ARTD-002.731.17...
PMP
Cleaning / Damaged paint
Clean the unit using the materials recommended in Instructions for use for MAMMOMAT
1000/3000 Nova (chapter Cleaning, Disinfection ) and repair any damaged paint.
System ON.
QSQ
Final test exposure
Take one final test exposure.
Siemens AG
Medical Solutions
Register 8
SPB7-230.101.03
Rev. 04
Page 1 of 2
07.03
CS PS 24
5 - 1
Opdimar
Maintenance Instructions