Service
To Gain access to the component side of the Front Panel Assembly
8-10
a. Remove the Front Panel Assembly from the instrument as mentioned previously.
b. Remove the 9/16 nut and washer securing the TEST OUT connector to the front panel.
c. Remove the 9 Pozidriv fixing screws that hold the circuit board in position on the front panel metalwork.
d. Carefully lift out the front panel circuit board and disconnect the Probe Power connector so that the
assembly may be lifted free.
Refitting the Front Panel assembly is the reverse of the above procedure.