Sony LMD-XH550MD Instructions For Use Manual page 22

Table of Contents

Advertisement

Be sure to connect equipment or cables
specified by Instructions for Use of this monitor
to the 12G-SDI/3G-SDI output connectors. If
you connect unspecified equipment or cables,
the monitor may affect the operation of the
connected equipment.
CLONE OUT connector (BNC type)
Output connector for serial digital component
signals of the recording equipment or monitor.
Outputs the displayed picture on the panel.
A signal is output from the CLONE OUT
connector only when the monitor is turned on.
When the monitor is turned off, the signal is
not output from the CLONE OUT connector.
Be sure to connect equipment or cables
specified by Instructions for Use of this monitor
to the CLONE OUT connector. If you connect
unspecified equipment or cables, the monitor
may affect the operation of the connected
equipment.
When using Clone Out in combination with the
recording device, connect the video output of
the recording device to the connector selected
in "AUX IN Setting." If the video output is
connected to another input connector, the
video loops when selecting the connector
connected to the recording device as a display
image on the monitor.
DVI-D input connector
Input connector for DVI Rev.1.0 applicable digital
signal.
The HDMI input connector, Display Port input
connector, and DVI-D input connector support
HDCP, a copy protection technology that
incorporates coding technology for digital video
signals.
SERIAL REMOTE (RS-232C) connector (D-
sub 9-pin, female)
Connect to the RS-232C control connector on the
external equipment connected to the monitor.
The monitor can be operated according to
control commands sent from connected external
equipment.
REMOTE connector (Stereo mini jack)
The monitor can be operated partially by
connecting the optional Foot Switch FS-24.
12V 2.5A (DC output) connector
Outputs 12 volts for connected external
equipment.
5V 2.0A (DC output) connector
Outputs 5 volts for connected external
equipment.
(Equipotential) terminal
Connects the equipotential plug.
 AC input connector
Connects the supplied AC power cord.
Warning
Using this unit for medical purposes
The connectors on this unit are not isolated.
Do not connect any device other than one which
conforms to IEC 60601-1 standards.
When an information technology device or AV
device that uses an alternating current is
connected, current leakage may result in an
electric shock to the patient or operator.
If use of such a device is unavoidable, isolate its
power supply by connecting an isolation
transformer, or by connecting an isolator
between the connecting cables.
After implementing these measures, confirm
that the reduced risk now conforms to IEC 60601-
1 standards.
22